Apply cotton wool pad with pressure to the customer’s finger
until the bleeding stops, discard the cotton wool pad as
clinical waste and offer a plaster.
•
Discard the capillary tube from the collection device and recap if
needed
•
Using the dedicated 32µl pipette and a clean dry tip,
press the plunger down to the first stop then lower the
tip into the blood sample
•
Once the tip is below the surface, gently release the
plunger fully
•
Check there are no bubbles in the tip
•
Hold the pipette and tip vertically over the sample pad on the test strip
•
Press the plunger down slowly, releasing the blood onto the sample pad
•
Dispose of the tip into the clinical waste container
Step 4
•
Open the Reflotron® measuring chamber lid
•
Immediately
insert the test strip into the Reflotron®
measuring chamber
•
When the strip is inserted correctly PLEASE CLOSE LID
will appear on the display. Close the lid.