4.
Electrical Connection
1)
Fix the motor, the customer has, to the base, if the customer purchased the pump without the motor.
2)
Remove the V-belt, when connecting the wires.
3) Switch in and watch the direction of rotation of the motor pulley. An arrow mark (like this:
⇒
) is
put on the model name plate to indicate the right direction of rotation. If the pump does not rotate
in the right direction, interchange any two of the three-phase leads.
Fig. 11 Diagram of motor connection and pulley rotation direction
4) Be sure to install a safety circuit such as a motor breaker or fuse in front of the motor, and an
electromagnetic switch (with thermal relay) for remote starting. Use the safety circuit capacity
shown in Table 4.
Attention: Be sure to use the motor breaker. Since a current 5 to 10 times the rated current flows
when the motor starts, a general-purpose breaker may trip at the motor starting current.
Table 4
Rated current for the standard motor
200V/50Hz 200V/60Hz 220V/60Hz
PVD-180
PVD-180B
0.4kW
2.11
2.00
1.95
PVD-360
PVD-360B
0.75kW
3.36
3.12
3.02
200V class Motor
Poles
:
4P
Model
Motor
Rated current
(
A
)
※
1
380V/50Hz 400V/50Hz 415V/50Hz 440V/60Hz 460V/60Hz
PVD-180
PVD-180B
0.4kW
1.13
1.14
1.15
1.03
1.04
PVD-360
PVD-360B
0.75kW
1.71
1.64
1.62
1.48
400V cLassMotor
Poles
:
4P
Model
Motor
Rated current
(
A
)
※
1
* 1 Rated current are about standard motors. Should you use other motor, confirm the specific by
the motor nameplate.