Chapter 7: Field Service Components
00.053.101, Revision C SonixMDP/SP/OP Service Manual
7-24
Table 7-1: SonixTOUCH System Case Connectivity Panels
1
PS2 Mouse Port
Not in use. (May or may not be present.)
2
PS2 Keyboard Port
Not in use. (May or may not be present.)
3
USB Ports (1 or 2)
Used by the wireless adapter (when applicable).
4
Serial Port
Not in use. (May or may not be present.)
OR
Used by the optional
ECG
module Serial connection (when applicable).
5
Parallel Port
Not in use. (May or may not be present.)
6
DVI Port
Connected to the DVI port on the Back Connectivity Panel (
7
USB Ports (2)
Used by the two (2) USB ports on the operator console and the USB connector from
the LCD display.
8
USB Ports (2)
Used by the two (2) USB ports on the Back Connectivity Panel (
).
9
Additional Audio Connections Not in use. (May or may not be present.)
10
Sound Connections
System Speaker connection (
green
).
Line-in (blue): may be used to connect an Ultrasonix-approved audio input device.
System Microphone connection: Disabled.
11
RS232 Serial Port
Used by the UPS.
12a
12b
RS232 Serial Port
Used by the operator console.
13
S-Video Output
(Back Connectivity
Panel
B
only)
When in use, connected to a video splitter module that outputs to a
non-USB Black & White (B&W) printer (using the
green
connector
and a BNC adapter).
Connect the printer to BNC (Output) 3 on Back Connectivity Panel
B
(
14
Video VGA Output
Not in use.
15
Digital Video DVI Video Output Used by the system's LCD display.
16
VGA Connector
Used by the operator console for video display.
17
Ethernet
For use by the Ethernet (LAN) connection on the Back Connectivity Panel. This port
supports 10/100Mb.
18
XTA Vdriver Serial Connector
Used by the serial ECG.
OR
Not in use. (May or may not be present.)
19
Console Power Connector
Used by the operator console.
20
Input (
1
) BNC
(Console
1
Button)
Connected to console
1
button, may be used to trigger a freeze or print function by
connecting directly to an Ultrasonix-approved switching device, such as a
Footswitch
.
During configuration, enable
Footswitch 1
for this connector. Refer to page
to configure the trigger action of the switching device.
Note:
For Back Connectivity Panel A only: If connection point
13
is active (factory-
installed option only), then
1
(BNC) will be configured for Output rather than Input
and a Footswitch cannot be installed.
21
Output (
2
) BNC
(Console
2
Button)
Connected to console
2
button, may be connected directly to an Ultrasonix-approved
triggered device.
22
IEEE 1394 Port
Not in use. (May or may not be present.)
23
S-Video Connector
Not in use. (May or may not be present.)
24
Video VGA Output
Not in use. (May or may not be present.)
25
Output (2) BNC
Not in use.
26
Input (2) BNC
Not in use.
28
DVI Digital Video Output
Used by the operator console.
Summary of Contents for SonixMDP
Page 1: ...SonixMDP SP OP Ultrasound System Service Manual...
Page 2: ......
Page 4: ......
Page 10: ...Table of Contents 00 053 101 Revision C SonixMDP SP OP Service Manual vi...
Page 12: ...List of Service Drawings 00 053 101 Revision C SonixMDP SP OP Service Manual ii...
Page 44: ...Chapter 4 Performance Testing 00 053 101 Revision C SonixMDP SP OP Service Manual 4 4...
Page 48: ...Chapter 5 Software 00 053 101 Revision C SonixMDP SP OP Service Manual 5 4...
Page 124: ...Chapter 9 Network Configuration 00 053 101 Revision C SonixMDP SP OP Service Manual 9 8...
Page 142: ...Chapter 11 Maintenance 00 053 101 Revision C SonixMDP SP OP Service Manual 11 16...
Page 150: ...Chapter 12 Troubleshooting Issues 00 053 101 Revision C SonixMDP SP OP Service Manual 12 8...
Page 166: ...Appendix C Ultrasonix Limited Warranty 00 053 101 Revision C SonixMDP SP OP Service Manual C 2...
Page 168: ...Appendix D Service Drawings 00 053 101 Revision C SonixMDP SP OP Service Manual D 2...