SonixMDP/SP/OP Service Manual 00.053.101, Revision C
Chapter 11: Maintenance
11-5
11.2.10 System Filter
This filter should be cleaned approximately every three (3) to six (6) months. Periodically, the system will present a
cleaning reminder message. Always clean the filter when this reminder is presented.
The SonixMDP/SP/OP comes equipped with a double layer filtering system. The first layer, or system filter, is
accessible without removing anything from the system – the filter can simply be pulled out. The second layer (fan
filtering) is attached to the modulo and necessitates modulo removal.
Cleaning the system filter every three (3) to six (6) months helps to safely prolong the modulo fan filter cleaning
interval (refer to
for the various maintenance procedure frequencies).
Use a vacuum cleaner to remove the dust from the system exhaust filters or vents. If a vacuum cleaner is not
available, dust off by hand or with an anti-static brush.
11.2.10.1Cleaning the System Filter
To Clean the System Filter:
1.
Power off and unplug the system.
2.
Taking care not to rip/damage the filter, g
ently
pull on the two (2) aluminum system filter frame handles until the
filter is free of its slot on the right side of the system.
3.
Vacuum thoroughly and
gently
reinstall the filter.
4.
Plug in and power on the system.
Warning:
Do not perform any internal system maintenance if the UPS breakers are turned on.
Caution:
Power off and unplug the system prior to cleaning the fan filters
Note:
For additional views, refer to Service Drawing:
System Filter Removal/Replacement
.
Caution:
When reinstalling, ensure the filter side faces toward the front of the system and the grid covering
the back of the filter faces towards the back.
Summary of Contents for SonixMDP
Page 1: ...SonixMDP SP OP Ultrasound System Service Manual...
Page 2: ......
Page 4: ......
Page 10: ...Table of Contents 00 053 101 Revision C SonixMDP SP OP Service Manual vi...
Page 12: ...List of Service Drawings 00 053 101 Revision C SonixMDP SP OP Service Manual ii...
Page 44: ...Chapter 4 Performance Testing 00 053 101 Revision C SonixMDP SP OP Service Manual 4 4...
Page 48: ...Chapter 5 Software 00 053 101 Revision C SonixMDP SP OP Service Manual 5 4...
Page 124: ...Chapter 9 Network Configuration 00 053 101 Revision C SonixMDP SP OP Service Manual 9 8...
Page 142: ...Chapter 11 Maintenance 00 053 101 Revision C SonixMDP SP OP Service Manual 11 16...
Page 150: ...Chapter 12 Troubleshooting Issues 00 053 101 Revision C SonixMDP SP OP Service Manual 12 8...
Page 166: ...Appendix C Ultrasonix Limited Warranty 00 053 101 Revision C SonixMDP SP OP Service Manual C 2...
Page 168: ...Appendix D Service Drawings 00 053 101 Revision C SonixMDP SP OP Service Manual D 2...