Chapter 12: Troubleshooting Issues
00.053.101, Revision C SonixMDP/SP/OP Service Manual
12-2
12.3 NO PRIMARY HARD DRIVE DETECTED
• Check the hard drive cable connections:
7.5 Front Block/Ultrasound Module Servicing
and
.
12.4 SYSTEM POWERS ON, LCD DISPLAY SHOWS MBR ERROR 1
Immediately after powering on, MBR 1 error is written to the LCD Display. This error occurs when the hard drive(s)
connection is loose or disconnected and the boot sequence is changed.
Check hard drive connections:
•
A.3 System Powers on, LCD Display shows MBR Error 1
12.5 SYSTEM SEEMS SLOW
System may be overheating:
• Check the CPU fan to ensure that it is working properly. If it is not, it will have to be replaced by a qualified Ultrasonix
Service Technician.
There may be issues with the RAM:
•
•
•
A software update may be required:
•
•
5.1.2 Software Update via a Downloaded File
DICOM may be turned on even though it is
not
configured. Check the following
DICOM
settings:
• A.4.3 Check Custom Keys Settings
• A.4.4 Check DICOM Settings
•
System may require data and file/folder cleanup:
•
7.1.3 Exporting Patient/Exam Data (As Required)
•
•
• defragment the hard drives.
System may be infected with a virus or spyware:
• perform a virus check.
Note:
This option applies only to software versions 5.4.3 and earlier.
Caution:
Once the Patient/Exam data has been successfully exported, delete the it from the system.
Store the Patient/Exam data in a secure place for future use/reference.
Summary of Contents for SonixMDP
Page 1: ...SonixMDP SP OP Ultrasound System Service Manual...
Page 2: ......
Page 4: ......
Page 10: ...Table of Contents 00 053 101 Revision C SonixMDP SP OP Service Manual vi...
Page 12: ...List of Service Drawings 00 053 101 Revision C SonixMDP SP OP Service Manual ii...
Page 44: ...Chapter 4 Performance Testing 00 053 101 Revision C SonixMDP SP OP Service Manual 4 4...
Page 48: ...Chapter 5 Software 00 053 101 Revision C SonixMDP SP OP Service Manual 5 4...
Page 124: ...Chapter 9 Network Configuration 00 053 101 Revision C SonixMDP SP OP Service Manual 9 8...
Page 142: ...Chapter 11 Maintenance 00 053 101 Revision C SonixMDP SP OP Service Manual 11 16...
Page 150: ...Chapter 12 Troubleshooting Issues 00 053 101 Revision C SonixMDP SP OP Service Manual 12 8...
Page 166: ...Appendix C Ultrasonix Limited Warranty 00 053 101 Revision C SonixMDP SP OP Service Manual C 2...
Page 168: ...Appendix D Service Drawings 00 053 101 Revision C SonixMDP SP OP Service Manual D 2...