![Ultrasonix SonixMDP Service Manual Download Page 59](http://html1.mh-extra.com/html/ultrasonix/sonixmdp/sonixmdp_service-manual_3519252059.webp)
SonixMDP/SP/OP Service Manual 00.053.101, Revision C
Chapter 7: Field Service Components
7-5
7.1.3
Exporting Patient/Exam Data (As Required)
Exam Import/Export
enables data to be copied to and from the system from within the system software, allowing
the creation of
Patient/Exam
data backups.
Equipment/Tools Required:
• USB media (memory stick, external hard drive, etc.).
To Export all Patient/Exam Data Using SonixMDP/SP/OP Software:
1.
Press the console button.
2.
Select
Import/Export
.
3.
Select all
Patients/Exams
.
4.
From the
Source
drop-down menu, select
Local Patient Data
.
5.
From the
Destination
drop-down menu, select the
Export Destination
.
6.
Select the
Export
button to begin the backup.
7.
When the
Export
is complete, the following message will be presented.
Note:
When creating a backup prior to servicing the modulo, be sure to select all
Patient/Exam
data currently
stored on the system.
If
Patient/Exam
data is exported using these instructions (
), it will not be necessary to use the
Backup
option (
Note:
Local Patient Data
is the only
Source
for exporting data.
Note:
The
Delete On Export Options
will not be available for selection until a valid
Destination
location is
selected.
To create a backup without removing the data from the system, leave both
Delete On Export Options
unchecked.
Note:
If
Delete On Export Options
were selected, the data will be deleted before the
Export
process is
complete message is presented.
Summary of Contents for SonixMDP
Page 1: ...SonixMDP SP OP Ultrasound System Service Manual...
Page 2: ......
Page 4: ......
Page 10: ...Table of Contents 00 053 101 Revision C SonixMDP SP OP Service Manual vi...
Page 12: ...List of Service Drawings 00 053 101 Revision C SonixMDP SP OP Service Manual ii...
Page 44: ...Chapter 4 Performance Testing 00 053 101 Revision C SonixMDP SP OP Service Manual 4 4...
Page 48: ...Chapter 5 Software 00 053 101 Revision C SonixMDP SP OP Service Manual 5 4...
Page 124: ...Chapter 9 Network Configuration 00 053 101 Revision C SonixMDP SP OP Service Manual 9 8...
Page 142: ...Chapter 11 Maintenance 00 053 101 Revision C SonixMDP SP OP Service Manual 11 16...
Page 150: ...Chapter 12 Troubleshooting Issues 00 053 101 Revision C SonixMDP SP OP Service Manual 12 8...
Page 166: ...Appendix C Ultrasonix Limited Warranty 00 053 101 Revision C SonixMDP SP OP Service Manual C 2...
Page 168: ...Appendix D Service Drawings 00 053 101 Revision C SonixMDP SP OP Service Manual D 2...