SonixMDP/SP/OP Service Manual 00.053.101, Revision C
Chapter 3: System Installation
3-17
3.3.3.2 Changing Fuses
Equipment/Tools Required:
• two (2) 3AG Slo-Blo, 7A/250V fuses
• fine tipped, narrow shafted, flat head screwdriver (e.g,
⅛
" (3mm) jewelers screwdriver).
To Change the Fuses:
1.
Power off and unplug the power cord from both the wall outlet and the system.
2.
Use the screwdriver to open the fuse box lid.
3.
Position the screwdriver at the top (as in
), using it to remove the fuse box.
4.
Replace one or both fuses, as required.
5.
Reinsert the fuse box.
6.
Close the fuse box lid.
Warning:
Do not perform any internal system maintenance if the UPS breakers are turned on.
Caution:
Ensure the correct voltage setting is visible and that it matches the setting on the EMI filter (
Summary of Contents for SonixMDP
Page 1: ...SonixMDP SP OP Ultrasound System Service Manual...
Page 2: ......
Page 4: ......
Page 10: ...Table of Contents 00 053 101 Revision C SonixMDP SP OP Service Manual vi...
Page 12: ...List of Service Drawings 00 053 101 Revision C SonixMDP SP OP Service Manual ii...
Page 44: ...Chapter 4 Performance Testing 00 053 101 Revision C SonixMDP SP OP Service Manual 4 4...
Page 48: ...Chapter 5 Software 00 053 101 Revision C SonixMDP SP OP Service Manual 5 4...
Page 124: ...Chapter 9 Network Configuration 00 053 101 Revision C SonixMDP SP OP Service Manual 9 8...
Page 142: ...Chapter 11 Maintenance 00 053 101 Revision C SonixMDP SP OP Service Manual 11 16...
Page 150: ...Chapter 12 Troubleshooting Issues 00 053 101 Revision C SonixMDP SP OP Service Manual 12 8...
Page 166: ...Appendix C Ultrasonix Limited Warranty 00 053 101 Revision C SonixMDP SP OP Service Manual C 2...
Page 168: ...Appendix D Service Drawings 00 053 101 Revision C SonixMDP SP OP Service Manual D 2...