Chapter 7: Field Service Components
00.053.205, Revision A SonixMDP/SP/OP Service Manual
7-36
7.7
TRANSDUCER TESTING
Ultrasound transducers are highly sensitive medical equipment requiring proper care and cleaning.
There are two possible sources when a client is experiencing image problems:
• transducer
• front block/ultrasound module's transducer MUX board.
After performing the following tests, report any failures to Ultrasonix Technical Support.
7.7.1
Testing the Transducer
Equipment/Tools Required:
• coin (with no rough edges).
To Test a Transducer:
1.
Connect the transducer to the system.
2.
Press the console
button and select the relevant transducer.
3.
Gently
slide the edge of any coin along the transducer face to test for a line(s) or shadow(s) in the image on the
LCD display.
4.
If there is a problem, unlock the transducer and move it to a different transducer port.
5.
Repeat
and check to see if the line/shadow(s) reappears in approximately the same place in the image
on the LCD display. If it does not disappear, then the transducer is the problem and it needs to be serviced.
Note:
for an example of an acceptable transducer image.
If there is any doubt about the image, contact Ultrasonix Technical Support and forward them a digital copy
of the image in question for verification of the diagnosis.
Summary of Contents for Sonix MDP
Page 1: ...Analogic Ultrasound Service Manual SonixMDP SP OPQ Ultrasound System...
Page 2: ......
Page 4: ......
Page 14: ...Chapter 2 System Specifications 00 053 205 Revision A SonixMDP SP OP Service Manual 2 2...
Page 46: ...Chapter 5 Software 00 053 205 Revision A SonixMDP SP OP Service Manual 5 4...
Page 120: ...Chapter 8 DICOM 00 053 205 Revision A SonixMDP SP OP Service Manual 8 24...
Page 147: ......