Chapter 7: Field Service Components
00.053.205, Revision A SonixMDP/SP/OP Service Manual
7-12
7.2.3
Turning ON the UPS Breakers
To Turn ON the UPS Breakers:
1.
Plug in the system.
2.
Turn
1: Battery Breaker
to the ON position.
3.
Turn
2: AC Input Breaker
to the ON position.
Warning:
Do not perform any internal system maintenance if the UPS breakers are turned on.
Note:
Breaker access is at the bottom of left side shroud (
).
Caution:
It is important to turn the breakers ON in the order specified.
Note:
Be sure to locate the breakers based on the system front (
3
).
1
1
2
2
3
Summary of Contents for Sonix MDP
Page 1: ...Analogic Ultrasound Service Manual SonixMDP SP OPQ Ultrasound System...
Page 2: ......
Page 4: ......
Page 14: ...Chapter 2 System Specifications 00 053 205 Revision A SonixMDP SP OP Service Manual 2 2...
Page 46: ...Chapter 5 Software 00 053 205 Revision A SonixMDP SP OP Service Manual 5 4...
Page 120: ...Chapter 8 DICOM 00 053 205 Revision A SonixMDP SP OP Service Manual 8 24...
Page 147: ......