SonixMDP/SP/OP Service Manual 00.053.205, Revision A
Chapter 7: Field Service Components
7-31
7.4.4
Reinstalling the Front Block/Ultrasound Module
Equipment/Tools Required:
• #2 Phillips screwdriver.
To Reinstall the Front Block/Ultrasound Module:
1.
Attach the new front block/ultrasound module using the #2 Phillips screwdriver to fasten the two hinges (two
screws per hinge).
2.
From the exterior of the modulo, using the #2 Phillips screwdriver, fasten the two screws holding the Side EMI
Strip over the door hinges.
Warning:
Do not perform any internal system maintenance if the UPS breakers are turned on.
Caution:
Always wear a grounding strap when opening and working inside the modulo.