Chapter 7: Field Service Components
00.053.205, Revision A SonixMDP/SP/OP Service Manual
7-6
7.1.4
Importing Patient/Exam Data (As Required)
Exam Import/Export
enables data to be copied to and from the system, allowing the reinstallation of
Patient/
Exam
data backups.
Equipment/Tools Required:
• USB media (memory stick, external hard drive, etc.) containing previously exported data.
To Import Patient/Exam Data:
1.
Press the console button.
2.
Select the
Import/Export
button.
3.
From the
Source
drop-down menu, select
the
Source of the data to be imported.
4.
From the list presented on the left hand side of the
Exam Import/Export
page, select all data for import.
5.
Select the
Import
button to begin the procedure.
6.
When the
Import
is complete, the following message will be presented.
Note:
When reinstalling
Patient/Exam
data after servicing the modulo, be sure to select all
Patient/Exam
data for
import to the system.
If the system boots properly with all
Patient/Exam
data intact, this procedure will not be required.
Note:
The following actions will auto-complete once the data
Source
is changed from
Local Patient Data
:
•
Destination
drop-down menu will change to
Local Patient Data
•
Export
button will change to
Import
.
Note:
If the data selected for
Import
is already available on the system, it will not be imported, i.e., it will not
overwrite the existing data.
Summary of Contents for Sonix MDP
Page 1: ...Analogic Ultrasound Service Manual SonixMDP SP OPQ Ultrasound System...
Page 2: ......
Page 4: ......
Page 14: ...Chapter 2 System Specifications 00 053 205 Revision A SonixMDP SP OP Service Manual 2 2...
Page 46: ...Chapter 5 Software 00 053 205 Revision A SonixMDP SP OP Service Manual 5 4...
Page 120: ...Chapter 8 DICOM 00 053 205 Revision A SonixMDP SP OP Service Manual 8 24...
Page 147: ......