![Ultrasonix Sonix MDP Service Manual Download Page 93](http://html1.mh-extra.com/html/ultrasonix/sonix-mdp/sonix-mdp_service-manual_3510469093.webp)
SonixMDP/SP/OP Service Manual 00.053.205, Revision A
Chapter 7: Field Service Components
7-37
7.7.2
Testing the Transducer Board
To Test the Transducer Board:
1.
Unlock and remove the transducer from its current transducer port.
2.
Re-attach and lock the transducer into a different transducer port.
3.
Press the console
button and select the relevant transducer.
4.
Test the transducer by taking a fresh image. If the line/shadow(s) in the image disappears, then the transducer
board is the problem and the front block/ultrasound module rather than the transducer needs to be serviced.
Figure 7-7: Acceptable Transducer Test Image
Note:
Refer to
for an example of an acceptable transducer image.
If there is any doubt about the image, contact Ultrasonix Technical Support and forward them a digital copy
of the image in question for verification of the diagnosis.
Close up of Near Field
Summary of Contents for Sonix MDP
Page 1: ...Analogic Ultrasound Service Manual SonixMDP SP OPQ Ultrasound System...
Page 2: ......
Page 4: ......
Page 14: ...Chapter 2 System Specifications 00 053 205 Revision A SonixMDP SP OP Service Manual 2 2...
Page 46: ...Chapter 5 Software 00 053 205 Revision A SonixMDP SP OP Service Manual 5 4...
Page 120: ...Chapter 8 DICOM 00 053 205 Revision A SonixMDP SP OP Service Manual 8 24...
Page 147: ......