UGO BASILE - I
TALY
Page 5
Series 58000 Instruction Manual (Rev.1)
In fact, the recorder status and functions are set-up automatically and visualized via the display.
The dedicated keys
RED
(DECOMPRESSION, which releases the pneumatic system and causes the
piston to speed back to its starting point, when depressurization is neces-
sary, see paragraph 2.4-Tail Cuffs).
BLUE
(START/REVERSE, see paragraph 7.3-Starting the Pressure Measure-
ment)
and the four functional keys
F1
-
F2
-
F3
-
F4
are the only operational commands.
The display and the functional keys combine to program the trial parameters selected by the op-
erator, to communicate the status of the experiment, any warning message, etc., see following
subheadings.
The upper rectangular window brings the linkage in sight and reassures the operator that the
pressure generator is doing its job.
2.3
Chart Recorder
This high-resolution, 50mm, real-time thermal printer of GSI LUMONICS, located on the
left side of the upper panel, has been selected as it offers a number of new features, in-
cluding warning messages to notify the user of undesirable conditions, thus preventing
errors.
See Figure 11 “Paper Roll”, self-explanatory on what concerns the replacement of the
chart roll.
The XE-50 communicates through a serial interface at 115,000 bits per second: when in
recorder mode it prints a simultaneous real-time waveform along with background grid,
text, and graphics.
The table below lists some specs. of this remarkable chart recorder:-
Printing method
:
direct, thermal
Print resolution
:
600 dots per 25 mm
Drive type
:
stepper motor
Chart specs.
:
2-inch thermal roll paper, 100 ft length
Chart speed
:
selectable
Chart skew
:
± 0.5 mm maximum over any 10 s interval, after first 150
Accuracy :
±
2%