Series 58000 Instruction Manual (Rev.1)
The datum (in our case the systolic pressure) can be
saved by the
F4
(
CON
) key or cancelled by the
F3
(
NUL
) key as shown in the display at this side.
In either case, the display shows the following picture,
which illustrates the diastolic pressure pattern (if previously
set).
The cursor mark is over the graph and the pressure figure
is below the base line.
As in the previous case, the cursor can be shifted to select
the best possible position.
In the case the Chart Recorder is on, the final picture of
the experiment is recorded (see the example taken from
a real experiment) where we read, from the top of the
data set:
⇒
date
⇒
time
⇒
systolic pressure (compression run)
⇒
diastolic pressure (compression run)
⇒
systolic pressure (decompression run)
2003/02/05
15:25:13
Systolic Up
121
Diastolic Up
070
Systolic Down 116
Diastolic Down 072
Pulse
Rate 545
Gender Male
Animal N° 12
⇒
diastolic pressure (decompression run)
⇒
pulse
rate
⇒
gender
⇒
animal
number
The latter two data, gender and animal number, are represented in the display as, e.g.,
M1 which means animal No. 1, male, F3 which means animal No. 3, female, etc.
8
CALIBRATION
To calibrate means to compare the pressure reading of 58500 with the data supplied by a
precision mercury pressure gauge (or any other certified pressure standard) capable to
cover the range 50 – 300 mm Hg.