Page 91
Message /
Symbol /
Problem
Failure Description
Corrective Action
205 kPa) for 2 seconds during
the sterilization stage.
Low
Pressure
(Cooling)
This message is displayed if
the pressure in chamber does
not reach the preset pressure
before initiating the cooling
stage.
Check and fix the
compressed air supply.
High
Pressure
This message is displayed if
Chamber Pressure raises 4.2
psi-29 kPa above sterilization
pressure (134°C = 304 kPa,
121°C = 205 kPa) for 2
seconds during the sterilization
stage.
Perform a new cycle.
High
Pressure
(Ending)
This message is displayed if
the system cannot reach
atmospheric pressure ± 10kPa
during the ending stage.
Perform a new cycle.
High
Pressure
(Exhaust)
This message is displayed if
the system cannot reach
preset pressure within 10
minutes from the beginning of
the exhaust stage.
Perform a new cycle.
Pressure
Time Error
This message is displayed if
the
system cannot reach the
required pressure conditions in
the chamber, after preset time,
during the air removal stage.
Verify that the autoclave
is not overloaded.
RTC Error -
Please Set
Current
Date and
Time
This message is displayed in
order to set the date and the
time.
Set Current Date And
Time.
If the problem persists,
call the technician.
Time Error
This message is displayed if
the real time clock is faulty.
Call the technician.
Door is
open
(During the
cycle)
This message is displayed
when the door is open: During
the cycle.
Close the door to
perform a new cycle.
Canceled
By User
This message is displayed
after the START/STOP key is
pressed and cycle aborted.
Wait until "cycle failed
–
canceled by user" or
"cycle end
– canceled by
user" is displayed.
Perform a new cycle.