Page 36
Nominal parameters default settings
•
Sterilization temperature: 121
C (250
F)
•
Sterilization time: 30 minutes
Operations Sequence
•
(PV, pre-vacuum models only)
Pulse low/Pulse high:
at one pulse it
will build vacuum down to 25 kPa.
•
Water Inlet:
Water enters the chamber
•
Heating:
The chamber is heated by actuation of electrical heaters
until the sterilization temperature is reached.
•
Sterilization:
Sterilization temperature is maintained constant during
the sterilization time.
•
Cooling:
N/A
•
Slow Exhaust:
Steam is exhausted from the chamber at a slow rate,
until it reaches the pressure of 30 kPa above the ambient pressure.
Then steam is exhausted out of the chamber at a fast rate until
pressure decreases to ambient pressure.
•
Drying:
N/A
•
Ending:
The system checks that End temperature and End pressure
have been reached, and then releases the door.
5.5
Program 5: Liquid A
–
Cooling (C
–
cooling models
only)
For liquids when the manufacturer recommends autoclaving at
temperatures of 121ºC (250
F). for 15 minutes.