Page 61
6.7.1
Reset atmospheric pressure
In this menu you can reset the atmospheric pressure value. To do so:
1.
Browse to the following folder:
Maintenance\Reset atmospheric pressure
The following screen will appear:
2.
Leave the door open for 2 minutes at least. Ambient temperature
should be less than 45°C.
Note:
Please reset the atmospheric pressure when you install the autoclave
for the first time, and each time you relocate or calibrate the
autoclave.
6.7.2
Printer test
I
n this menu you can check the normal function of the printer. The
printer will print the list of errors.
Browse to the following folder:
Maintenance\Printer test
The following screen will appear to confirm that the test has been done.