No Water
This message is displayed if
the electrode in the chamber
did not sense water within the
preset time.
1.
check and fix the
mineral free water
supply.
2.check and clean the
water inlet filter.
3.
Clean the water
level electrode.
Analog
Input Error
This message is displayed
when any Temperature sensor
or Pressure sensor is
disconnected or out of range.
Call for service.
Chamber
temperatur
e not in
range
This message is displayed if
the temperature in the
chamber is too high or too low
from the normal range.
Call for service.
Chamber
pressure
not in range
This message is displayed if
the pressure in the chamber is
too high or too low from the
normal range.
Call for service.
I/O Card
Failed
This message is displayed if
I/O card is faulty (both while
cycle is running or not).
Call for service.
I/O card is
not
connected
This message is displayed if
I/O card is disconnected (both
while cycle is running or not).
Call for service.