Page 45
Put one PT100 inside the bottle, leave the second one hanging in the
chamber outside the bottle (see below).
Nominal parameters default settings
•
Heating temperature (default): 80
C (176
F).
•
Heating time: 20 minutes.
Operations sequence:
•
Water Inlet:
Water enters the chamber
•
Heating:
The chamber is warmed up until the Keep-Heat
temperature is reached inside the chamber.
•
Keep Heat:
Keep-Heat temperature is maintained constant for the
preset Keep-Heat time.
•
Cooling:
N/A.
•
Fast Exhaust:
the steam is exhausted out of the chamber at a fast
rate until pressure decreases to ambient pressure.
•
Drying:
N/A
•
Ending:
The system checks that End temperature and End pressure
have been reached, and then releases the door.
Note:
Some water may remain in the chamber.