Page 63
Cycle parameters\Drying\Dry Time
2.
Change the parameter as desired.
Drying stage is divided into two stages. For each stage you can set
the total time, on time and off time. So during the Dry First Stage, the
heating elements are on for the time set in Dry Heat On 1, then they
go off for the time set in Dry Heat Off 1, and then this sequence is
repeated during the entire Dry First Stage Time. The same is true for
the Dry Second Stage.
6.8.2
Add Dry Time
This subdirectory allows you to change the Dry Time parameter for
the current cycle.
Browse to the following folder:
Cycle parameters\Drying\Add Dry Time
Change the parameter as desired