Page 92
Message /
Symbol /
Problem
Failure Description
Corrective Action
Cycle
Failed
This message and symbol are
displayed if an error occurs
before sterilization cycle is
completed.
Perform a new cycle.
Air Error
This message is displayed at
the end of the cycle If the
autoclave does not reach the
atmospheric pressure after 10
minutes.
Wait until the autoclave
reaches the atmospheric
pressure and perform a
new cycle.
Compresse
d air supply
error
This message is displayed in
case of a compressed air
supply malfunction.
Check and fix the
compressed air supply.
Periodical
check time
exceeded -
Please call
for service
The periodical maintenance
time has passed.
Call for service.
"Supplied
Distilled
Water
Error"
Mineral free water reservoir
empty"
Fill the mineral free
water reservoir.
Cycle
counter
exceeded -
Please call
for service
Number of cycles, since last
periodical maintenance,
exceeded the preset number
as defined by "cycle counter"
parameter.
Call for service.
Power
Down
This message is displayed if
power down has occurred
during the cycle. (this
message will print out in the
printer after the autoclave will
turn on).
Turn on the autoclave
and wait until the
autoclave is ready
(reaches the safe
condition) and perform a
new cycle.
Supply
distilled
water error
(digitat
input
option)
This message is displayed in
case of a mineral free water
supply malfunction.
Check and fix the
mineral free water
supply
Supply
water error
(digitat
input
option)
This message is displayed in
case of a city (tap) water
supply malfunction.
Check and fix the city
(tap) water supply.
Compresse
d air supply
error
This message is displayed in
case of a compressed air
supply malfunction.
Check and fix the air
supply.