Page 29
Cycle by
clock
This message is displayed if
the user presses
START/STOP
key while the
"cycle by clock" mode is
active.
Enter the Admin menu as
described in this manual to
change the time or to cancel this
option.
Atmospheri
c pressure
not set
This message id displayed
in order to set the
atmosphere pressure by
opening the door for 5
minutes.
Open the door for 5 minutes in
order to set the Atmospheric
pressure.
Critical
settings
have been
updated,
Please
restart
machine in
order for
changes to
be updated
If a change of the autoclave
setting was made, a restart
operation is required.
Restart the autoclave in order
for changes to be updated.
This message is displayed if
the electrode in the
chamber senses water.
Perform a new cycle to drain the
chamber.
Supplied
Distilled
Water Error
Mineral free water reservoir
empty"
Fill the mineral free water
reservoir.