APPENDIX D: TROUBLESHOOTING
TD1000C™Oil in Water Monitor
30
TD-1000C Oil In Water Monitor
Operation Manual
PN 101687
REV G
5. From the above screen, press <
ENT
>
OPER: xxxx
HRS
LED: ON FLOW: ON
Oper:
Indicates how many hours the monitor has been in operation since installation.
LED:
Indicates whether the LED light source is operating properly.
Flow:
Indicates whether the flow is ON/OFF.
6. From the above screen, press <
ENT
>
REF:
x.xxxV
REFCAL:
xxx.xV
Ref:
This value is the current reference measurement.
Ref Cal:
This value is the measurement taken during the last cell calibration procedure.
7. From the above screen, press <
ENT
>
FL:
xx.xxV
FL
: This value is the current raw, non-calibrated fluorescence measurement from the detector.
8. From the above screen, press <
ENT
>
CAL TEMP:
xx.xC
MAX TEMP:
xx.xC
Cal Temp:
This is the temperature of the calibration solution during the calibration procedure.
Max Temp:
This is the maximum process water temperature the instrument has measured since
the original installation.
9. From the above screen, press <
ENT
>
BLANK: x.x
CAL STD: xxx.x
Blank:
This is the stored “blank” fluorescence value taken during the calibration procedure.
Cal Std:
This is the stored “Cal Std” taken during the calibration procedure.
For definitions, ranges, and default values of these items, see
Appendix F
.
D.2 Service Assistance
The experienced technical staff of Turner Designs Hydrocarbon Instruments is available to assist you in
troubleshooting the TD1000C™ instrument.
Turner Designs Hydrocarbon Instruments
Telephone USA: +1 559 253 1414
E-mail: [email protected].
E-mail: [email protected]
Hours: 8:00 am - 5:00 pm Pacific Standard Time