TD1000C™Oil in Water Monitor
CALIBRATION
PN 101687
REV G
TD-1000C Oil In Water Monitor
Operation Manual
17
STEP 5
Press <
ENT
>:
CALIBRATE SYSTEM
<1> TO START
STEP 6
Press <
1
>:
BLANK
SOLUTION
Use the syringe for the blank solution to flush the flow cell with 60 mL of blank solution. Then inject
another 60 mL blank solution and allow it to remain in the flow cell by leaving the syringe attached.
Bubbles trapped during injection are a possible source of instrument error. With the syringe in a
vertical position, tap against a solid object to move bubbles to the needle end of the syringe. Then,
force bubbles out by pushing a small amount of solution through the needle end of the syringe.
STEP 7
Press <
ENT
>:
BLANK
xx.x
Any value less than 25 is acceptable. If blank is greater than 25 and <
Ý
> is pressed, an error message
will be received. Press <ESC> to abort calibration and check the blank solution. Press <
Ý
> when the
reading is stable. The monitor will display a flashing WAIT/wait message in the lower right-hand corner
of the screen while the monitor registers the blank.
Then, the screen will display:
TO CONTINUE
PRESS <ENT>
STEP 8
Press <
ENT
>:
CALIBRATION
SOLUTION
Use a clean syringe to flush the flow cell with 60 mL of calibration solution via the luer-lock Injection
Port. Then inject another 60 mL calibration solution and allow it to remain in the flow cell by leaving the
syringe attached.