APPENDIX G: SOFTWARE/KEYPAD FUNCTIONS
TD1000C™Oil in Water Monitor
44
TD-1000C Oil In Water Monitor
Operation Manual
PN 101687
REV G
G.2.7
Diskette Key
The <
Í
> key causes the message “CONNECT TO SERIAL PORT” to be displayed on the LCD. The
key has no other functionality but reminds the user that if he wishes to electronically log data, he must
connect the monitor to a computer via the internal serial port.
Í
G.3
User Identification
To change System Values or to calibrate the instrument, a four-digit User ID (0258) is required. For
security, a Master ID, different from the User ID, is required to view or change the User ID. To change
the User ID, contact your Turner Designs Hydrocarbon Instruments representative or Turner Designs
Hydrocarbon Instruments Service Department at +1 559 253 1414 for assistance with the Master ID.
Once an ID has been entered, if the keypad is not used for 15 minutes, the monitor will automatically
return to the
HOME
screen. The ID will have to be entered again before the monitor values can be
changed or calibration can be performed.
G.4 Fluorescence Display
After the monitor is powered-up or after calibration, the measurement displayed will not react
immediately, but will respond after a delay of about 10 seconds.
G.5
LCD Contrast
The contrast of the Liquid Crystal Display (LCD) can be adjusted on any screen (except during
calibration) by pressing <
ñ
> or <
ò
>. <
ñ
> makes the LCD lighter and <
ò
> makes the LCD darker.