CALIBRATION
TD1000C™Oil in Water Monitor
18
TD-1000C Oil In Water Monitor
Operation Manual
PN 101687
REV G
STEP 9
Press <
ENT
>:
CAL SOLUTION
xx
Wait for the reading to stabilize.
When the CAL SOLUTION reading is stable and between 5 and 10, press <Ý>. The monitor will display
a WAIT/wait message in the lower right-hand corner of the screen while it registers the calibration
solution.
Then, the screen will display:
TO CONTINUE
PRESS <ENT>
STEP 10
Remove the syringe and press <
ENT
>:
PRESS <1> TO END
CALIBRATION
Press <
1
> to accept the calibration settings. The calibration is now complete.
You must press <1> or the calibration will revert to the previous settings!
STEP 11
Open the flow cell inlet valve. The valve is open when the yellow valve handle is oriented vertically.
Shift the flow cell outlet valve to the process flow position. The outlet valve is correctly positioned when
the green handle is vertical.
If your instrument has PVC wetted parts, position the inlet valve handle vertically and the outlet valve in
the process position, with the end of the valve handle pointed down. See Appendix K.
STEP 12
Press <
HOME
> to return to the
HOME
Screen.
STEP 13
Establish a new offset value as described in
Appendix F
.
5.3 NOTES ON CALIBRATING
1. Calibrate TD1000C™ monitor when you have time to go through all steps without interruption
(
≈
10 minutes). If the keypad is not used for 15 minutes, the monitor will automatically return
to the HOME screen. The previous calibration will be maintained.
2. AC Relay output will be OFF during calibration.