CALIBRATION
TD1000C™Oil in Water Monitor
16
TD-1000C Oil In Water Monitor
Operation Manual
PN 101687
REV G
5.2 FLUORESCENCE CALIBRATION PROCEDURE
The TD1000C™ is factory calibrated for Number 6 Fuel Oil applications. Fluorescence Calibration is
not required during initial start-up. If your application is a hydrocarbon other than Number 6 Fuel Oil,
the factory calibration is also used, but only the calibration solution value is changed.
The procedure detailed in the following paragraphs should be used for periodic fluorescence calibration
requirements during operation.
All calibration steps must be completed for changes to be registered!
Refer to
Figures 1 through 5 for locations of items. A diagram of display screens is located in
Appendix
F
.
STEP 1
Close the flow cell inlet valve. The valve is closed when the yellow valve handle is oriented horizontally.
Shift the flow-cell outlet valve to the drain position. The valve is correctly positioned when the green
valve handle is oriented horizontally.
If your instrument has PVC wetted parts, position the inlet valve handle horizontally and the outlet valve
in the drain position, with the end of the valve handle pointed up. See Appendix K.
STEP 2
Clean the monitor by injecting cleaning solution with a syringe into the flow cell. The syringe screws
onto the Luer-Lock fitting on the inlet of the flow cell (see
Figure 1
). Fill syringe with 60 mL of cleaning
solution. Screw syringe onto fitting. Inject cleaning solution at a slow, steady rate into flow cell and
allow it to remain there for 3–5 minutes. Use a clean 60 mL syringe to flush the flow cell with 60 mL of
blank solution.
A syringe is provided in the TD1000C™ Calibration Accessory Kit (Turner Designs Hydrocarbon
Instruments Part No 101665).
STEP 3
Press <
ENT
> from the
HOME
screen:
1. CALIB 2. DATA
CAL xx DAY(S) AGO
STEP 4
Press <1>. The monitor will display a request for ID entry (unless recently entered).
PLEASE INPUT ID:
Enter valid USER ID (originally 0258) on the keypad if requested.