TD1000C™Oil in Water Monitor
START-UP
PN 101687
REV G
TD-1000C Oil In Water Monitor
Operation Manual
9
Remove the monitor enclosure face carefully. Failure to do so may result in breakage of the ribbon
cable attached to the back of the enclosure face.
The ribbon cable is indexed on the PCB to insure proper alignment upon reassembly. There are two
terminal strips within the instrument. Terminals #1–6 are for AC connections and terminals #A–K are for
input/output connections.
To connect a wire to the terminal strip
1. Turn off main power to the instrument at breaker.
2. Loosen the proper terminal screw (use screwdriver provided) and insert wire from below into
terminal.
3.
Tighten screw firmly. A termination legend is provided on the backside of the enclosure door
for reference.
To disconnect a wire from the terminal strip
1. Turn off main power to the instrument at breaker.
2. Loosen the termination screw and pull the wire out of the terminal.
When finished, carefully insert the ribbon cable onto the PCB and replace enclosure face with the 6
screws.
Signal wires and power wiring should not be run in the same conduit. Failure to separate or shield
these wires will result in electrical interference.
Improper installation, use, application or unauthorized servicing of this equipment voids all
guarantees.
Turner Designs Hydrocarbon Instruments recommends installing a secondary surge arrestor, or
lightning suppressor, at the main power supply to the TD1000C™, according to the manufacturer’s
instructions.
4 START-UP
Before start-up, the following items (not supplied with the TD1000C™) should be obtained:
•
De-Ionized (DI) or Distilled Water.
•
10% muriatic (hydrochloric) acid solution for cleaning scaling and iron oxide.
•
Liquid detergent for cleaning hydrocarbon fouling.
If necessary, please contact your Turner Designs Hydrocarbon Instruments representative or contact
Turner Designs Hydrocarbon Instruments at +1 559 253 1414 for additional information.