Operation
35
Note
: In this example, the program asks you to confirm
the choice of Calibration Factor, as “0.25”. This
value is within the allowable range, but is
considered an unusually low value. For this
reason, the software asks for a confirmation
before proceeding.
Note
: This action only adds a new Cal Factor to a list of
available factors. This does
not
program this new
Cal Factor into the instrument. To reprogram the
instrument, continue with the procedure.
4. After entering the new Cal Factor into the list of available
factors, the instrument is programmed with this new
factor. Highlight the selected cal factor (in this example,
“coal”), and click one of the two “arrow” buttons. In this
example, the “coal” Cal Factor was programmed into Cal
Fac
tor #4, replacing the previous selection of “oil mist.”