Setting-Up
11
!
W A R N I N G
Use of USB connection to be used only in an area known
to be nonhazardous.
The power supply cannot run the SidePak
monitor and charge the
battery at the same time. Make sure the instrument is turned off
to initiate the charging cycle. If the display does not show
CHARGING BATTERY, it is
not
charging.
When the power supply is first plugged into the AM510, the
display will first show the message DETECTING BATTERY
TYPE. If any of the TSI NiMH battery packs are detected, the
display will show CHARGING BATTERY. Once the battery is fully
charged, the display will show CHARGING COMPLETE. The
instrument may be turned on at any time during or after the
charging process by pressing the PAGE key. Turning the
instrument on during the charging process will abort charging.
If the AA battery pack is detected (regardless of installed cell
type) or there is no battery pack at all, the display will show
CHARGING OFF after 30 seconds. Press the PAGE key to turn
the instrument on or off at any time.