Operation
25
Prog 1 and Prog 2 Mode Protocols
Percent
memory
required.
Displays the percent of SidePak memory
required to perform a Prog 1 or a Prog 2
mode sample. To store the results of a
Prog 1 or Prog 2 mode sample, the
Available Memory
must be equal to or
greater than the
Percent memory
required
.
3.
Enter
the following for Prog 1 and Prog 2 modes:
Start Date
Start Time
Enter the date and time to begin the
sample:
If you enter a blank for a start date, the
sample begins whenever the specified
start time occurs.
If you enter a blank for the start time,
both start date and start time are
ignored, and the sample begins when
the operator manually starts the
sample.
Log interval
Enter the log interval to use for the test.
Test length
Enter the length for the sample:
If you enter a value, the instrument
automatically turns off when the last
test is complete.
If you enter a blank, you must
manually stop the sample.
Number of
tests
Enter the number of tests to perform.
These tests will happen sequentially, with
the
Time between tests
spaced in
between each repetition.
Time between
tests
If you have specified more than one
Number of tests
, enter the time between
tests. If you enter 0 or blank, the next test
is started immediately after the last test is
complete.
While you are entering values for Prog 1 and Prog 2
modes, the
Percent Memory Required
is dynamically
updated to show the amount of SidePak memory
required to take the programmed sample. If the protocol
you have defined requires more than 100% of memory
(or more than the
available
memory), you can decrease
Summary of Contents for SIDEPAK AM510
Page 1: ...SIDEPAK PERSONAL AEROSOL MONITOR MODEL AM510 USER GUIDE ...
Page 2: ......
Page 66: ...Appendix B 58 This page intentionally left blank ...
Page 69: ...61 Appendix D CSA Certificate of Compliance ...
Page 70: ...Appendix D 62 ...
Page 71: ...CSA Certificate of Compliance 63 ...
Page 72: ...Appendix D 64 This page intentionally left blank ...