Chapter 3
28
6. You can now disconnect your SidePak
monitor from the
computer.
After programming the SidePak monitor with TrakPro Data
Analysis Software, attach the SidePak monitor to the worker
(or location) and turn it on. Put it into Prog 1 or Prog 2 mode
(whichever you want) using the keypad.
1. From SURVEY MODE, press PAGE to access MAIN
MENU.
2. Scroll to Data Log and press
.
3. Scroll to Run Prog 1 (or Run Prog 2) and press
.
4. If desired, lock the keypad to prevent tampering. Press
and hold
the ▲ key, then press
. Repeat to unlock.
Things You Should Know About Using Prog 1 and Prog 2
Modes
If you press the PAGE
or
key during programmed
operation, the instrument responds with a message,
asking you to confirm your choice to end the program
(unless the keypad is locked).
If the programmed start date has already passed, the
program will not execute. The instrument responds with a
message indicating
INVALID PRGM; START DATE
.
It is always best to enter both a start time and start date!
However, setting the start time but no start date causes
the SidePak monitor to start at the specified time
regardless of the date. The instrument may not shut
down prior to beginning logging.
If no start time is set, the SidePak monitor begins
sampling immediately after the program is selected and
the
key is pressed.
If no test length is set, the SidePak monitor takes
samples continuously until the program is deliberately
terminated, using the
key, or until memory is full.
When a preprogrammed test ends, the SidePak monitor
automatically shuts off.
Summary of Contents for SIDEPAK AM510
Page 1: ...SIDEPAK PERSONAL AEROSOL MONITOR MODEL AM510 USER GUIDE ...
Page 2: ......
Page 66: ...Appendix B 58 This page intentionally left blank ...
Page 69: ...61 Appendix D CSA Certificate of Compliance ...
Page 70: ...Appendix D 62 ...
Page 71: ...CSA Certificate of Compliance 63 ...
Page 72: ...Appendix D 64 This page intentionally left blank ...