31
(without black mark) is placed on the black mark sensor.
If yes, please follow the following steps to calibrate sensor sensitivity
agian.
1.Press the button
under “
3
” icon to return to Bline Sensor menu.
2. Press the button
under “
]” icon to enter “Auto Bline” menu.
3. After insert paper on the black mark sensor, press the button
under “V” icon to continue.
4. Press the button
under “+” icon to increase Stock AD value by
50. And press the button
under “V” icon to calibrate the sensor
sensitivity.
4.1.2.2.2 Manual Bline:
If Auto Bline sensor calibration successfully, it’s not necessary to
calibrate the sensor manually. This function is used to fine tune sensor
sensitivity. When enter the [Manual Bline] option, you will see following
message.
Stock AD on label
Stock AD on black
Press the
button under “+” and “-“ to increase and decrease
sensor intension, and press the
under “V” to set the value to
printer
When adjust the sensor sensitivity manually, we can get the different
Stock AD on black mark and paper in each sensor intension. And the
difference between these two values must be larger than 50 levels then
printer can register the label correctly.
4.1.2.3 Continuous:
FAILED
Stock AD = 190
3
Bline Calibrate
240
+ - V
Gap Sensor
Stock AD = 153
Intension = 212
+ - V
Gap Sensor
Stock AD = 249
Intension = 212
+ - V