Handle the strip with clean and dry hands; lancing site for the test   

    should be cleaned and dried thoroughly. 

 

 

Do not perform the test while a cell phone is used beside; prevent 

    the electronic signal interference.

Oxygen concentration in air is low may cause the test inaccurate;

    there is no effect from altitude up to 10,000 feet above sea level.

Medical therapy for intravenous injection, such as immunoglobulin 

    preparations and peritoneal dialysis solutions

,

containing sugar 

    compounds may cause higher test results

.

Hematocrit 

(

Hct

)

 range from 30-55% is acceptable

.

Hct less than

 

  

30% may cause elevated result

.

Hct above 55% may cause reduced 

  

glucose results versus a whole blood reference

Do not drop the product

.

Keep the lancets and lancing device out of the reach of the children.

 

o

Keep the meter and test strips at room temperature (10-30 C

).

1.3 Intended Use

TBG-460GC  Blood Glucose and Cholesterol Monitoring System 
is intended to be used for quantitative measurement of blood glucose 
level and cholesterol level in fresh capillary whole blood drawn from the
fingertips. It is intended for use outside of body (In Vitro diagnostic
use). It should be use only as an aid to monitor blood glucose and 
cholesterol level by individuals at home and by healthcare professionals
at clinical settings.

Summary of Contents for TBG-460GC

Page 1: ...Blood Glucose Cholesterol 2 in 1 Monitoring System User s Manual 2012 03 12 Rev 00 Model TBG 460GC...

Page 2: ...unit Chapter 3 How to perform Test 3 1 Before Testing 3 2 Perform the Blood Glucose Test 3 4 Special messages 3 5 Care after the test Chapter 4 Chapter 5 Meter Memory Function Chapter 6 Installing Rep...

Page 3: ...ation procedures with illustrated directions and help you find the required informations Please contact the local distributor for questions and enquiries WARNING Please carefully read the instructions...

Page 4: ...55 is acceptable Hct less than 30 may cause elevated result Hct above 55 may cause reduced glucose results versus a whole blood reference Do not drop the product Keep the lancets and lancing device o...

Page 5: ...companying documents Storage temperature limitation Use by EU Representative Catalogue number Product number Manufactured by 0123 1 5 System Components Test Strip Port Insert test strip here The meter...

Page 6: ...At least 10uL of blood sample is essential Reaction Zone Reaction zone of Total Cholesterol CHOL Glucose Test Strip Blue Color Sample Inlet The inlet where the whole blood or control solution is drawn...

Page 7: ...s are listed as follows Set the Year 1 When the meter enter setting mode the first item on the screen such as 2012 will blink on screen which means Year 2 Press the up down button to 3 After you set t...

Page 8: ...up dowm button until the correct measurement unit appears 2 Press the memory button will show OFF on the screen and power off Chapter 3 How to perform Test 3 1 Before Testing 7 Coding the Meter TBG 4...

Page 9: ...so the same as marked on the test vial Remove the code strip should appear on the screen indicating that the meter has been successfully coded Cautions If E E appears on the screen during coding proce...

Page 10: ...remove the protective lancet cover 3 Replace the end cap by snapping it back onto the lancing device body 4 Select the skin penetration depth by turning the depth selector until the desired number can...

Page 11: ...luence the result lancing 4 Touch the sample Inlet of the strip to the drop of blood sample until the meter sounds a beep which means the sample has drawn into strip successfully 5 The screen will dis...

Page 12: ...s 3 5 Care after the Test Directions 1 Improper usage of the meter such as dropping bumping or other violent impact may cause the meter damaged 2 Please keep meter away by interferences of magnetic el...

Page 13: ...eep the strips in dry condition 2 Be careful to make sure the test mode and code number on the screen are the same as the one marked on the test strip vial If not remove the test strip and re code the...

Page 14: ...mode only total cholesterol test mode do not have the day average function Chapter 5 Meter Memory Function Please read this user s manual for control solution test carefully before use Do not reuse t...

Page 15: ...l hear a beep sound which indicates that meter is 3 Place back the cover onto the meter ready 5 After the 28day average the latest stored result M01 appears on the screen Then you can read the stored...

Page 16: ...bent hole Revlew the instructiona and try again with a new test strip If the problem persists please contact the dealer for service E 9 Chapter 7 Storage and Handling of Meter and Strips Due to sensit...

Page 17: ...ter and the test strip have reached at temperature within the operating range E U It could be caused by a used test strip or a temporary or permanent electronic problem Repeat the test with a new test...

Page 18: ...se level 75 mg dL 4 17mmol L 20 at Total Cholesterol 150mg dL 3 88mmol L Electrochemical Biosensor Plasma equivalent Glucose 20 600 mg dL 1 1 33 3 mmol L Total Cholesterol 100 400 mg dL 2 59 10 35 mmo...

Page 19: ......

Reviews: