-3-
-
Handle the strip with clean and dry hands; lancing site for the test
should be cleaned and dried thoroughly.
-
Do not perform the test while a cell phone is used beside; prevent
the electronic signal interference.
-
Oxygen concentration in air is low may cause the test inaccurate;
there is no effect from altitude up to 10,000 feet above sea level.
-
Medical therapy for intravenous injection, such as immunoglobulin
preparations and peritoneal dialysis solutions
,
containing sugar
compounds may cause higher test results
.
-
Hematocrit
(
Hct
)
range from 30-55% is acceptable
.
Hct less than
30% may cause elevated result
.
Hct above 55% may cause reduced
glucose results versus a whole blood reference
.
-
Do not drop the product
.
-
Keep the lancets and lancing device out of the reach of the children.
o
-
Keep the meter and test strips at room temperature (10-30 C
).
1.3 Intended Use
TBG-460GC Blood Glucose and Cholesterol Monitoring System
is intended to be used for quantitative measurement of blood glucose
level and cholesterol level in fresh capillary whole blood drawn from the
fingertips. It is intended for use outside of body (In Vitro diagnostic
use). It should be use only as an aid to monitor blood glucose and
cholesterol level by individuals at home and by healthcare professionals
at clinical settings.