Truly TBG-460GC User Manual Download Page 18

-17-

Weight

Dimensions                   

Battery Life                    

Battery Type                 

Memory                      

Relative Humidity          

Operation Temperature 

Strip Storage Condition

Hematocrit Range        

Sample Volume              

Measuring Range         

Measuring Time              About 10 seconds (Glucose),

26 seconds (Total Cholesterol)

Test Sample                      Fresh Capillary whole blood

Calibration                    

Principle                           

System Accuracy           ± 20% at Glucose level 

75 mg/dL (4.17mmol/L) 

± 20% at Total Cholesterol 

150mg/dL(3.88mmol/L)

Electrochemical Biosensor 

Plasma-equivalent

Glucose 20 - 600 mg/dL (1.1 - 33.3 mmol/L)
Total Cholesterol 100 - 400 mg/dL (2.59 - 10.35 mmol/L)

About 1 - 1.5 µL (Glucose)
About 10 µL (Total Cholesterol)

30 - 55% (for Glucose)
35 - 50% (for Total Cholesterol) 

10 - 30ºC (50 - 86ºF)

10 - 40ºC  (50 - 104ºF)

Less than 95%

360 for Glucose, 50 for Total Cholesterol

One 3V (CR2032) lithium battery

Approximately 1,000 tests

82*54*21.5mm (Length, Width, Height)

About 45 g (with battery)

Chapter 9 Specification 

Summary of Contents for TBG-460GC

Page 1: ...Blood Glucose Cholesterol 2 in 1 Monitoring System User s Manual 2012 03 12 Rev 00 Model TBG 460GC...

Page 2: ...unit Chapter 3 How to perform Test 3 1 Before Testing 3 2 Perform the Blood Glucose Test 3 4 Special messages 3 5 Care after the test Chapter 4 Chapter 5 Meter Memory Function Chapter 6 Installing Rep...

Page 3: ...ation procedures with illustrated directions and help you find the required informations Please contact the local distributor for questions and enquiries WARNING Please carefully read the instructions...

Page 4: ...55 is acceptable Hct less than 30 may cause elevated result Hct above 55 may cause reduced glucose results versus a whole blood reference Do not drop the product Keep the lancets and lancing device o...

Page 5: ...companying documents Storage temperature limitation Use by EU Representative Catalogue number Product number Manufactured by 0123 1 5 System Components Test Strip Port Insert test strip here The meter...

Page 6: ...At least 10uL of blood sample is essential Reaction Zone Reaction zone of Total Cholesterol CHOL Glucose Test Strip Blue Color Sample Inlet The inlet where the whole blood or control solution is drawn...

Page 7: ...s are listed as follows Set the Year 1 When the meter enter setting mode the first item on the screen such as 2012 will blink on screen which means Year 2 Press the up down button to 3 After you set t...

Page 8: ...up dowm button until the correct measurement unit appears 2 Press the memory button will show OFF on the screen and power off Chapter 3 How to perform Test 3 1 Before Testing 7 Coding the Meter TBG 4...

Page 9: ...so the same as marked on the test vial Remove the code strip should appear on the screen indicating that the meter has been successfully coded Cautions If E E appears on the screen during coding proce...

Page 10: ...remove the protective lancet cover 3 Replace the end cap by snapping it back onto the lancing device body 4 Select the skin penetration depth by turning the depth selector until the desired number can...

Page 11: ...luence the result lancing 4 Touch the sample Inlet of the strip to the drop of blood sample until the meter sounds a beep which means the sample has drawn into strip successfully 5 The screen will dis...

Page 12: ...s 3 5 Care after the Test Directions 1 Improper usage of the meter such as dropping bumping or other violent impact may cause the meter damaged 2 Please keep meter away by interferences of magnetic el...

Page 13: ...eep the strips in dry condition 2 Be careful to make sure the test mode and code number on the screen are the same as the one marked on the test strip vial If not remove the test strip and re code the...

Page 14: ...mode only total cholesterol test mode do not have the day average function Chapter 5 Meter Memory Function Please read this user s manual for control solution test carefully before use Do not reuse t...

Page 15: ...l hear a beep sound which indicates that meter is 3 Place back the cover onto the meter ready 5 After the 28day average the latest stored result M01 appears on the screen Then you can read the stored...

Page 16: ...bent hole Revlew the instructiona and try again with a new test strip If the problem persists please contact the dealer for service E 9 Chapter 7 Storage and Handling of Meter and Strips Due to sensit...

Page 17: ...ter and the test strip have reached at temperature within the operating range E U It could be caused by a used test strip or a temporary or permanent electronic problem Repeat the test with a new test...

Page 18: ...se level 75 mg dL 4 17mmol L 20 at Total Cholesterol 150mg dL 3 88mmol L Electrochemical Biosensor Plasma equivalent Glucose 20 600 mg dL 1 1 33 3 mmol L Total Cholesterol 100 400 mg dL 2 59 10 35 mmo...

Page 19: ......

Reviews: