4–2
CALIBRATION SCHEDULE
The Troxler Model 5850v2 Gyratory Compactor is calibrated at the
factory, and requires no initial calibration upon receipt.
Troxler recommends the following calibration schedule for the
Model 5850v2. If calibrating the compaction pressure and specimen
height, calibrate the pressure first and then calibrate the height.
NOTE
To ensure proper compactor operation, calibrate the
compaction pressure and specimen height after
changing the ram pressure (see page 3–7).
PRESSURE CALIBRATION
See page 4–6 for details on pressure calibration.
Calibrate the compaction pressure every 3 months.
SPECIMEN HEIGHT CALIBRATION
See page 4–10 for details on specimen height calibration.
Calibrate the specimen height daily.
Calibrate the specimen height after calibrating the compaction
pressure.
ANGLE VERIFICATION AND OFFSET
See page 4–3 for details on angle verification.
Verify the angle as described on page 4–3 after every 800 samples
or once per year, whichever comes first. If the angle is more than
±0.03° from the target, use the
Angle Offset
function as described on
page 4–12.
Summary of Contents for 5850v2
Page 10: ...x LIST OF TABLES Table Title Page Table 2 1 Control Panel Keys and Button 2 2...
Page 17: ...Model 5850v2 1 7 1 INTRODUCTION Figure 1 1 Model 5850v2 Gyratory Compactor...
Page 27: ...Model 5850v2 2 3 2 SETUP OPERATION Figure 2 1 Model 5850v2 Control Panel Layout...
Page 35: ...Model 5850v2 2 11 2 SETUP OPERATION part numbers 106952 or 106953...
Page 37: ...Model 5850v2 2 13 2 SETUP OPERATION the chamber door...
Page 62: ...4 8 Figure 4 3 Routing Load Cell Cables ROUTE CABLES THROUGH NOTCH LOAD CELL...
Page 68: ...4 14 NOTES...
Page 76: ...5 8 NOTES...
Page 86: ...Appendix A 10 NOTES...
Page 99: ...Model 5850v2 Index...
Page 100: ...NOTES...
Page 102: ...12...