2–14
COMPACTING THE SPECIMEN
The Model 5850v2 Gyratory Compactor provides a fully automatic
method of controlling compaction. In
Automatic
mode, the operator
begins the compaction cycle using a single keystroke. The Model
5850v2 compacts the specimen based on the number of gyrations or
the target specimen height specified by the operator. In
Automatic
mode, the compactor can also automatically print or download
compaction data upon completion of the compaction cycle.
Place the loaded mold into the compaction chamber and close the
chamber door. Press
START
.
If there is no mold in the chamber, or if the mold is not in the proper
location, the compactor displays an error message.
If the mold is in place, the compactor begins the compaction cycle.
The compactor clamps the mold, raises the ram into position,
induces the angle of gyration, and begins compaction. During the
compaction cycle, the compactor displays a screen similar to the
one shown below:
-Gyrating-
Gyration: #:##
Spec. Ht.: ###.# mm
G:100 AI:1.16 P:600
Where:
Gyration #
=
Number of gyrations completed
Spec. Ht.
=
The current specimen height
G
=
Operator-selected number of gyrations (this value
is 999 if the compactor is set to compact to a
target specimen height)
AI
=
Internal angle of gyration
P
=
Operator-selected compaction pressure
Summary of Contents for 5850v2
Page 10: ...x LIST OF TABLES Table Title Page Table 2 1 Control Panel Keys and Button 2 2...
Page 17: ...Model 5850v2 1 7 1 INTRODUCTION Figure 1 1 Model 5850v2 Gyratory Compactor...
Page 27: ...Model 5850v2 2 3 2 SETUP OPERATION Figure 2 1 Model 5850v2 Control Panel Layout...
Page 35: ...Model 5850v2 2 11 2 SETUP OPERATION part numbers 106952 or 106953...
Page 37: ...Model 5850v2 2 13 2 SETUP OPERATION the chamber door...
Page 62: ...4 8 Figure 4 3 Routing Load Cell Cables ROUTE CABLES THROUGH NOTCH LOAD CELL...
Page 68: ...4 14 NOTES...
Page 76: ...5 8 NOTES...
Page 86: ...Appendix A 10 NOTES...
Page 99: ...Model 5850v2 Index...
Page 100: ...NOTES...
Page 102: ...12...