Model 5850v2
3–9
3.
SE
TU
P M
ENU
Select
3
to change Post Gyrate Delay time.
NOTE
Gyration process will proceed as normal.
Input Number of
Minutes
XX
Press <ENTER>
Valid entries of
1
to
90
minutes for extra time when post delay is
needed.
-Post Gyrate Delay-
MM:SS
Remaining
<ESC> to Abort
When the post-gyrate delay is enabled, after gyrations are complete
the screen will display:
NOTE
Where MM represents minutes and SS represents seconds.
During the delay, the pump remains on. The carriage plate is
centered, the pressure is set to minimum and the ram position
remains constant. The time counts down until it has elapsed or
pressing
ESC
to Abort
early.
Once the time has expired, the process is completed as usual; ram is
lowered, pump is turned off and data is output (
enabled
).
Summary of Contents for 5850v2