3–4
CHANGE SETTINGS
The
Change Settings
selection from the
Setup
menu allows
you to:
Select whether the compactor will run for a specified number of
gyrations or until the specimen has been compacted to a
specified height
Specify the number of gyrations (from 1 to 999) the compactor
will run when compacting a specimen
Set the desired specimen height
Set the desired gyration angle (0.00 to 3.0 degrees)
Define the desired ram pressure (200 to 1000 kPa)
Select the mold diameter (150-mm, 100-mm, or 4-inch)
To access these options, press
2
from the
Setup
menu (see page
3–2). The
Change Settings
menu displays, as shown below. Use
the arrow keys to scroll through the menu options. To select an
option, press the numeric key that corresponds to desired option.
The following sections describe each selection.
-Change Settings-
1- Operation Mode
2- # of Gyrations
3- Height
-Change Settings-
4- Angle
5- Pressure
6- Mold Diameter
-Change Settings-
7- Post Gyrate Delay
Summary of Contents for 5850v2
Page 10: ...x LIST OF TABLES Table Title Page Table 2 1 Control Panel Keys and Button 2 2...
Page 17: ...Model 5850v2 1 7 1 INTRODUCTION Figure 1 1 Model 5850v2 Gyratory Compactor...
Page 27: ...Model 5850v2 2 3 2 SETUP OPERATION Figure 2 1 Model 5850v2 Control Panel Layout...
Page 35: ...Model 5850v2 2 11 2 SETUP OPERATION part numbers 106952 or 106953...
Page 37: ...Model 5850v2 2 13 2 SETUP OPERATION the chamber door...
Page 62: ...4 8 Figure 4 3 Routing Load Cell Cables ROUTE CABLES THROUGH NOTCH LOAD CELL...
Page 68: ...4 14 NOTES...
Page 76: ...5 8 NOTES...
Page 86: ...Appendix A 10 NOTES...
Page 99: ...Model 5850v2 Index...
Page 100: ...NOTES...
Page 102: ...12...