1–12
ASSEMBLY
1
1
.
.
The hydraulic reservoir is filled with a premium quality, light
grade (ISO VG 32, SAE 10) hydraulic oil prior to shipping.
Before using the compactor, check the oil level as follows:
a
a
.
.
Remove the rear service panel from the compactor.
b
b
.
.
Locate the fill level sight glass as shown in Figure 1–2.
NOTE
Use only a light grade (ISO VG 32, SAE 10, Troxler part
number 018203) hydraulic oil to fill the hydraulic
reservoir.
2
2
.
.
Ensure that the oil level is at the 30-degree mark on the sight
glass, as shown in Figure 1–3. If oil is needed:
a
a
.
.
Remove the reservoir fill cap (see Figure 1–2) and place a
funnel into the fill tube as shown in Figure 1–4.
b
b
.
.
Add oil until the level is at the 30-degree mark.
c
c
.
.
Remove the funnel and replace the reservoir fill cap.
d
d
.
.
Re-install the rear service panel.
Figure 1
–2. Hydraulic Reservoir
RESERVOIR
FILL CAP
FILL LEVEL
SIGHT GLASS
Summary of Contents for 5850v2
Page 10: ...x LIST OF TABLES Table Title Page Table 2 1 Control Panel Keys and Button 2 2...
Page 17: ...Model 5850v2 1 7 1 INTRODUCTION Figure 1 1 Model 5850v2 Gyratory Compactor...
Page 27: ...Model 5850v2 2 3 2 SETUP OPERATION Figure 2 1 Model 5850v2 Control Panel Layout...
Page 35: ...Model 5850v2 2 11 2 SETUP OPERATION part numbers 106952 or 106953...
Page 37: ...Model 5850v2 2 13 2 SETUP OPERATION the chamber door...
Page 62: ...4 8 Figure 4 3 Routing Load Cell Cables ROUTE CABLES THROUGH NOTCH LOAD CELL...
Page 68: ...4 14 NOTES...
Page 76: ...5 8 NOTES...
Page 86: ...Appendix A 10 NOTES...
Page 99: ...Model 5850v2 Index...
Page 100: ...NOTES...
Page 102: ...12...