Model 5850v2
4–9
4.
CA
LIBRA
TIO
N
CAUTION
Use caution when opening and closing the chamber door
to prevent catching or pinching the load cell cables.
5
5
.
.
Press
ENTER
. The compactor measures the unloaded
pressure on the load cell. During this measurement, the control
unit displays:
Pressure Cal/Verify
Testing Calibration
Load Cell
Please Wait
NOTE
If the load cell is not properly connected, the control
unit briefly displays the following error message. Check
the load cell connection. Press
ENTER
to continue.
Pressure Cal/Verify
Communication Error
Check Load Cell
Press <ENTER>
6
6
.
.
After measuring the unloaded pressure on the load cell, the
compactor raises the ram until the load cell contacts the upper
puck plate. The compactor then applies pressure and displays:
Pressure Cal/Verify
Applying Pressure
During calibration and verification, the compactor raises the
compaction ram and performs a series of pressure measurements at
high and low pressure. Upon completion, the compactor verifies the
pressure calibration, and then displays the compaction pressure as
measured internally by the compactor (
System
) and as measured by
the load cell (
Load Cell
):
Summary of Contents for 5850v2
Page 10: ...x LIST OF TABLES Table Title Page Table 2 1 Control Panel Keys and Button 2 2...
Page 17: ...Model 5850v2 1 7 1 INTRODUCTION Figure 1 1 Model 5850v2 Gyratory Compactor...
Page 27: ...Model 5850v2 2 3 2 SETUP OPERATION Figure 2 1 Model 5850v2 Control Panel Layout...
Page 35: ...Model 5850v2 2 11 2 SETUP OPERATION part numbers 106952 or 106953...
Page 37: ...Model 5850v2 2 13 2 SETUP OPERATION the chamber door...
Page 62: ...4 8 Figure 4 3 Routing Load Cell Cables ROUTE CABLES THROUGH NOTCH LOAD CELL...
Page 68: ...4 14 NOTES...
Page 76: ...5 8 NOTES...
Page 86: ...Appendix A 10 NOTES...
Page 99: ...Model 5850v2 Index...
Page 100: ...NOTES...
Page 102: ...12...