4-8
To recalibrate the gauge for the densities outside the factory
calibration range, press
〈
ON/YES
〉
. To disable the
Special
Calibration
feature, press
〈
OFF/NO
〉
at the above display and
〈
ON/YES
〉
at the disable inquiry. After disabling this feature, the
gauge will return to the
Ready
mode.
To enter a known B value obtained with the procedure available
from Troxler, press
〈
ON/YES
〉
. To have the gauge calculate the
recalibration, press
〈
OFF/NO
〉
. If entering a new known B value,
see the following explanation. For gauge-calculated special
calibration, see page 4-9.
ENTERING A NEW B VALUE
To change the value of the flashing digit for the measurement depth,
use the up and down arrows. To accept the flashing value and select
the next digit, press
〈
START/ENTER
〉
.
The gauge displays the current B value. To change the value of the
flashing digit, use the up and down arrows. To accept the flashing
digit and select the next digit, press
〈
START/ENTER
〉
.
SPECIAL CALIB.
Enter B Value?
Depth = xx in
(
↑
↓
or ENTER)
B VAL = x.xxxx
(
↑
↓
or ENTER)
Summary of Contents for 3430-M
Page 9: ......
Page 14: ...xii NOTES...
Page 19: ...Model 3430 1 5 Figure 1 1 Model 3430 Gauge and Accessories INTRODUCTION...
Page 22: ...1 8 NOTES...
Page 26: ...2 4 Figure 2 3 Backscatter Surface Density Effects Top Layer Effect Curves...
Page 28: ...2 6 Figure 2 4 Effect of Moisture on Depth of Measurement...
Page 48: ...3 20 NOTES...
Page 62: ...4 14 NOTES...
Page 66: ...5 20 D xxxx M xxxx Use keys Moist R xxxx to view data Moist Avg xxxx to view data...
Page 88: ...A 14 NOTES...
Page 111: ...Model 3430 C 17 Figure C 2 3430 Gauge Assembly MAINTENANCE...
Page 115: ...Model 3430 C 21 Figure C 5 3430 Gauge Preamplifier Assembly MAINTENANCE...
Page 117: ...Model 3430 C 23 Figure C 6 3430 Gauge Scaler Assembly MAINTENANCE...
Page 128: ...E 2 Date MS DS Date MS DS STANDARD COUNT LOG Gauge Serial Number...
Page 129: ...Model 3430 E 3 Date MS DS Date MS DS STANDARD COUNT LOG Gauge Serial Number STANDARD COUNT LOG...
Page 130: ...E 4 Date MS DS Date MS DS STANDARD COUNT LOG Gauge Serial Number...
Page 140: ...Index 8 NOTES...