4-6
TRENCH OFFSETS
If the Model 3430 gauge is to be used for moisture or density
measurements in a trench or within two feet (0.6 m) of a large
vertical structure, a trench offset may be required. If used, the trench
offset adjusts all moisture measurements but only the density
measurements from backscatter through four inches (10 cm).
To perform a trench offset:
✓
Take the daily standard count (outside the trench) and record the
density count (
DS
) and moisture count (
MS
)values.
✓
Place the gauge on the reference standard block in the trench the
same distance from the wall as the anticipated readings.
Do not
take another standard count.
✓
Set the count time to four minutes.
✓
With the source rod in the
SAFE
(standard count) position, take
a four-minute count. To start the count, press the
〈
START/ENTER
〉
key.
✓
Record the trench density count (
DC
Trench
) and moisture count
(
MC
Trench
.).
✓
Subtract the daily standard count values from the trench count
values:
Dens Const = (DC
Trench
) – DS
Moist Const = (MC
Trench
) – MS
To enable a trench offset, choose
Trench
from the
Offset
options
under the
Special
functions. The gauge requests the
Dens Const
and
Moist Const
values determined above. The procedure for entering
the values is the same as for moisture and density offsets,
ignoring
the
±
sign on the display.
Summary of Contents for 3430-M
Page 9: ......
Page 14: ...xii NOTES...
Page 19: ...Model 3430 1 5 Figure 1 1 Model 3430 Gauge and Accessories INTRODUCTION...
Page 22: ...1 8 NOTES...
Page 26: ...2 4 Figure 2 3 Backscatter Surface Density Effects Top Layer Effect Curves...
Page 28: ...2 6 Figure 2 4 Effect of Moisture on Depth of Measurement...
Page 48: ...3 20 NOTES...
Page 62: ...4 14 NOTES...
Page 66: ...5 20 D xxxx M xxxx Use keys Moist R xxxx to view data Moist Avg xxxx to view data...
Page 88: ...A 14 NOTES...
Page 111: ...Model 3430 C 17 Figure C 2 3430 Gauge Assembly MAINTENANCE...
Page 115: ...Model 3430 C 21 Figure C 5 3430 Gauge Preamplifier Assembly MAINTENANCE...
Page 117: ...Model 3430 C 23 Figure C 6 3430 Gauge Scaler Assembly MAINTENANCE...
Page 128: ...E 2 Date MS DS Date MS DS STANDARD COUNT LOG Gauge Serial Number...
Page 129: ...Model 3430 E 3 Date MS DS Date MS DS STANDARD COUNT LOG Gauge Serial Number STANDARD COUNT LOG...
Page 130: ...E 4 Date MS DS Date MS DS STANDARD COUNT LOG Gauge Serial Number...
Page 140: ...Index 8 NOTES...