A-12
EMERGENCY PROCEDURES
If the nuclear gauge is lost or stolen, then immediately notify the
gauge owner’s Radiation Safety Officer (RSO).
The gauge owner should complete the emergency contact
information on the lines furnished below. (Note that company
refers to the gauge owner’s company, not Troxler Electronic
Laboratories.) This information should be readily available to
the gauge operator at all times.
If a gauge is damaged, then follow the steps below:
✓
Locate the gauge and/or source.
✓
Do not touch or move the gauge.
✓
Immediately cordon off an area around the nuclear gauge and/or
source. A radius of fifteen feet (5 m) will be sufficient. Do not
leave the area unattended.
✓
Keep all unauthorized personnel from the nuclear gauge.
✓
If a vehicle is involved, it must be stopped until the extent of
contamination, if any, can be established.
✓
The gauge operator should perform a visual inspection of the
nuclear gauge to determine if the source housing and/or
shielding has been damaged.
✓
Use a survey meter to measure the dose rate at a distance of
three feet (1 m) from the gauge.
The company RSO is
Call the RSO at
The regulatory agency is
Call the agency at
Summary of Contents for 3430-M
Page 9: ......
Page 14: ...xii NOTES...
Page 19: ...Model 3430 1 5 Figure 1 1 Model 3430 Gauge and Accessories INTRODUCTION...
Page 22: ...1 8 NOTES...
Page 26: ...2 4 Figure 2 3 Backscatter Surface Density Effects Top Layer Effect Curves...
Page 28: ...2 6 Figure 2 4 Effect of Moisture on Depth of Measurement...
Page 48: ...3 20 NOTES...
Page 62: ...4 14 NOTES...
Page 66: ...5 20 D xxxx M xxxx Use keys Moist R xxxx to view data Moist Avg xxxx to view data...
Page 88: ...A 14 NOTES...
Page 111: ...Model 3430 C 17 Figure C 2 3430 Gauge Assembly MAINTENANCE...
Page 115: ...Model 3430 C 21 Figure C 5 3430 Gauge Preamplifier Assembly MAINTENANCE...
Page 117: ...Model 3430 C 23 Figure C 6 3430 Gauge Scaler Assembly MAINTENANCE...
Page 128: ...E 2 Date MS DS Date MS DS STANDARD COUNT LOG Gauge Serial Number...
Page 129: ...Model 3430 E 3 Date MS DS Date MS DS STANDARD COUNT LOG Gauge Serial Number STANDARD COUNT LOG...
Page 130: ...E 4 Date MS DS Date MS DS STANDARD COUNT LOG Gauge Serial Number...
Page 140: ...Index 8 NOTES...