5-10
RECALL
The
Recall
function allows the operator to view the data from the
last
reading. Even though the 3430 gauge does not store multiple
readings, this function displays the latest data.
To access the
Special
functions, press
〈
SPECIAL
〉
.
To access the
Recall
feature, press
〈
START/ENTER
〉
.
The gauge displays the data from the last measurement. Scroll
through the screens using the up and down arrow keys.
OFFSET
For information on offsetting gauge readings, refer to Chapter 4.
STAT TEST
The
statistical stability test
, or
stat test
, may be performed to
validate the normal operation of the gauge. Erratic readings or
readings that seem to fluctuate may indicate a problem with the
gauge. In the event the readings are suspect, a stat test may be
executed.
A
stat test
consists of twenty 1-minute counts. After the twenty
counts, the gauge calculates the standard deviation. This standard
deviation is compared to a theoretical standard deviation value.
Ideally this ratio should be one. However, the 3430 gauge pre-scales
(or divides) the counts by 16, resulting in an ideal ratio of 0.25. The
acceptable limits for the ratio are from 0.17 to 0.33. The gauge is
considered to be unstable if the ratio is outside these limits.
Summary of Contents for 3430-M
Page 9: ......
Page 14: ...xii NOTES...
Page 19: ...Model 3430 1 5 Figure 1 1 Model 3430 Gauge and Accessories INTRODUCTION...
Page 22: ...1 8 NOTES...
Page 26: ...2 4 Figure 2 3 Backscatter Surface Density Effects Top Layer Effect Curves...
Page 28: ...2 6 Figure 2 4 Effect of Moisture on Depth of Measurement...
Page 48: ...3 20 NOTES...
Page 62: ...4 14 NOTES...
Page 66: ...5 20 D xxxx M xxxx Use keys Moist R xxxx to view data Moist Avg xxxx to view data...
Page 88: ...A 14 NOTES...
Page 111: ...Model 3430 C 17 Figure C 2 3430 Gauge Assembly MAINTENANCE...
Page 115: ...Model 3430 C 21 Figure C 5 3430 Gauge Preamplifier Assembly MAINTENANCE...
Page 117: ...Model 3430 C 23 Figure C 6 3430 Gauge Scaler Assembly MAINTENANCE...
Page 128: ...E 2 Date MS DS Date MS DS STANDARD COUNT LOG Gauge Serial Number...
Page 129: ...Model 3430 E 3 Date MS DS Date MS DS STANDARD COUNT LOG Gauge Serial Number STANDARD COUNT LOG...
Page 130: ...E 4 Date MS DS Date MS DS STANDARD COUNT LOG Gauge Serial Number...
Page 140: ...Index 8 NOTES...