3-16
where:
WD
= Wet density in kg/m
3
or pcf
DD
= Dry density in kg/m
3
or pcf
%PR
= Percent Proctor (This value is valid only if an
appropriate target has been entered for the material
being tested.)
MOIST
= Moisture value in kg/m
3
or pcf
% MOIST
= Percent moisture
Air Void
= See description below
Void Ratio
= See description below
MOIST CR
= Moisture count ratio
DENS. CR
= Density count ratio
M Count
= Moisture counts as read by the gauge
D Count
= Density counts as read by the gauge
xxxxx PCF
xx.x %
DD:
%PR:
xxxxx PCF
xx.x %
Moist:
% Moist:
xx.x%
Air Void:
Void Ratio
MOIST CR: xx.x
DENS CR: xx.x
M Count: xxxxx
D Count: xxxxx
(Use
↑
&
↓
keys)
PCF
WD:
xxxxx
Summary of Contents for 3430-M
Page 9: ......
Page 14: ...xii NOTES...
Page 19: ...Model 3430 1 5 Figure 1 1 Model 3430 Gauge and Accessories INTRODUCTION...
Page 22: ...1 8 NOTES...
Page 26: ...2 4 Figure 2 3 Backscatter Surface Density Effects Top Layer Effect Curves...
Page 28: ...2 6 Figure 2 4 Effect of Moisture on Depth of Measurement...
Page 48: ...3 20 NOTES...
Page 62: ...4 14 NOTES...
Page 66: ...5 20 D xxxx M xxxx Use keys Moist R xxxx to view data Moist Avg xxxx to view data...
Page 88: ...A 14 NOTES...
Page 111: ...Model 3430 C 17 Figure C 2 3430 Gauge Assembly MAINTENANCE...
Page 115: ...Model 3430 C 21 Figure C 5 3430 Gauge Preamplifier Assembly MAINTENANCE...
Page 117: ...Model 3430 C 23 Figure C 6 3430 Gauge Scaler Assembly MAINTENANCE...
Page 128: ...E 2 Date MS DS Date MS DS STANDARD COUNT LOG Gauge Serial Number...
Page 129: ...Model 3430 E 3 Date MS DS Date MS DS STANDARD COUNT LOG Gauge Serial Number STANDARD COUNT LOG...
Page 130: ...E 4 Date MS DS Date MS DS STANDARD COUNT LOG Gauge Serial Number...
Page 140: ...Index 8 NOTES...