5.1.4 Warm-up time from a cold start
The detector will be fully operational within typically 30 minutes after being powered up with
performance levels guaranteed after a maximum of 240 minutes. It is the System responsibility to
check for this delay.
5.2 INTEGRATION
5.2.1 Front
protective
label
The detector is delivered with a large label attached to the input carbon fiber composite substrate. It
provides marks to align the X-ray beam collimation with the detector sensitive array.
! ! ! CAUTION ! ! !
Remove the protective label before operation (or before detector integration on the
System).
! ! ! ATTENTION ! ! !
Oter l’étiquette de protection avant la mise en service ou l’intégration du détecteur
5.2.2 Mechanical fixations :
In all cases, the detector must be fixed to the System by means of all eight screws M4 (refer to
outline drawing for location), with a threaded length of 4 mm minimum and a total length of 7 mm.
For accessory fixation on the detector back side (such as heat sinks), use up to eight screws M4
(refer to outline drawing for location), with a length of 6 mm maximum.
Refer to appendix C (outline drawing) for the location of these fixations.
The detector can be operated in any position in space (0 to 360° in any X, Y, Z directions).
5.3 X-RAY GENERATOR CONNECTION
The PIXIUM 4600 provides three different possibilities to enable X-Radiation. Depending on the
System receiving the detector, X-ray emission can be enabled using :
•
Current-loop signal linked with Com signals on the mini-DIN connector and designed to
drive an optocoupler device
•
Microcontroler messages on the RS-232 serial link
•
Taxi
messages linked with image data on the optic fiber