background image

14 

 

SYSTEM CALIBRATION 

 

Calibration Overview

 

 
Digital system calibration is accomplished in two steps: zero calibration (F16) and positive span 
calibration (F17). A third calibration for negative span calibration is also available via F15. You 
can calibrate up to 7 points per side.  
 
Here is the recommended sequence for multiple fixtures and multiple calibration points: 

1.  Put load cell onto test fixture  
2.  Go to F17 and press the down key; indicator prompts for the first calibration point  
3.  key in number 111111 and press NET/GROSS key ; indicator will show “FIT” mo-

mentarily and then automatically record the fixture reference point 

4.  Follow F17 procedure as written for up to seven calibration points  (do not exit F17) 
5.  Repeat steps 2 thru 6 for negative span calibration (F15) 
6.  Put the Loadcell onto the final fixture 
7.  Go to F16 and perform zero calibration  

 
For dual wireless A/D modules systems (e.g. TI-500 RFTM-2BE), a digital corner calibration fea-
ture is also available. It does not require a specific test weight value, but the maximum weight 
that should be used is approximately 25% of the rated capacity of the platform. 
 

NOTE

: Please perform corner calibration prior to executing zero/span calibration. 

Digital Corner Calibration (dual wireless units only) 

 

1.  Switch off the force data instrument by pressing and holding the ON/PRINT key for about 5 

seconds. 

2.  Press and hold down the ON/PRINT key (about 20 seconds) until the screen shows “-F-”. 

3.  Press the ON/PRINT key until the screen displays “-B-”. 

4.  Scroll down once using the ZERO/OFF key to enter the “Bluetooth” menu. Instrument shows  

“B  1”. 

5.  While  in  the  Bluetooth  Setup  mode,  scroll to  "

B  10

",  and then  scroll  down once  using  the  

ZERO/OFF key to enter corner calibration menu. The instrument will display a value. This 
value is the internal A/D count and can prove useful when trying to troubleshoot setup prob-
lems. 

6.  Remove all items from the weighing platform. 

7.  Press the NET/GROSS key to save the zero-point value. The display will show "

Corn-1

".  

8.  Place the test weight on the load cell/corner #1 of the weighing platform. 

9.  Wait 2-3 seconds to allow the internal reading to stabilize. 

10.  Press the NET/GROSS key to save the load cell/corner #1 calibration. The display will show 

"

Corn-2

". 

11.  Repeat steps 8-10 for the remaining load cells/corners. At the conclusion of corner #4 cali-

bration, the display will show "

Set--

". 

12.  Press the NET/GROSS key to finish the corner calibration and revert back up to B10. 

13. 

In  order  to  save  all  parameter  settings,  it  is  imperative  to  exit  all  menus  in  the  following 
manner:  At  the  parameter  level,  press  the  UNITS  (up)  key  twice  (two  times)  to  exit  the  
Bluetooth Menu. The instrument displays ‘SAvE’ and then automatically powers off.

 

Summary of Contents for TI-500 FDI Series

Page 1: ...MODEL TI 500 FDI Force Data Instrument Installer s Manual Revision 1 1 October 14 2016...

Page 2: ...OF TI 500 FORCE DATA INSTRUMENT 2 GETTING STARTED 3 SYSTEM CONFIGURATION 4 CONFIGURATION MENUS 4 ENTERING THE SETUP F CONFIGURATION MENU 5 SETUP F MENU DESCRIPTIONS 5 ENTERING THE USER A OR COM C MENU...

Page 3: ...libration of the force data instrument For operation and troubleshooting please refer to the separate user s guide Installation of TI 500 force data instrument Find a suitable location for the instrum...

Page 4: ...batteries to operate not included To install the batteries remove the plastic battery cover from the rear panel Observe proper direction polari ty of the batteries Replace the battery cover The TI 500...

Page 5: ...NOTE This menu will not appear for cabled systems see F29 The configuration menus are laid out in the following vertical arrangement Top Menu selection level Parameter level Selection level or functi...

Page 6: ...meter settings it is imperative to exit all menus in the following manner At the parameter level press the UNITS up key twice two times to exit the Setup Menu The instrument displays SAvE and then aut...

Page 7: ...re Press ZERO OFF key to begin se quence F16 Zero Calibra tion Places instrument into the zero calibration routine Scrolling down with the ZERO OFF key one level begins the procedure Press ZERO OFF ke...

Page 8: ...e AdC Internal A D cabled 1rAdIo One external wireless A D module 2rAdIo Two external wireless A D modules AdC 1rAdIo 2rAdIo F30 Special Appli cation Used to select one special application feature sub...

Page 9: ...o go from A1 to A2 press the ON PRINT key To go from A2 back to A1 press the TARE key 6 Once you have arrived at the proper A menu parameter e g A 1 press the ZERO OFF down key once to arrive at the s...

Page 10: ...bit 7_o_2 7 data bits with odd parity bit and two stop bits 8_n_1 8_n_2 7_n_1 7_n_2 7_E_1 7_E_2 7_o_1 7_o_2 A3 C3 Serial Port Mode Selects the mode of the serial port Refer to Appendix B for more info...

Page 11: ...lects the printed format for date USA mm dd yy IntL dd mm yy USA IntL A20 Set System Time Date Allows you to set the system time and date Pressing the ZERO OFF key to scroll down one level begins the...

Page 12: ...ashing digit 10 After entering the exact value press the NET GROSS key to save the value The screen displays dA_xx where xx is the current day of the month e g 14 One digit will be flash ing 11 Use th...

Page 13: ...uously out both serial ports TEST1 on COM1 and TEST2 on COM2 End test manually by pressing the NET GROSS Set key A24 U6 Keyboard Test Displays a keycode for each key pressed on the keypad See Table be...

Page 14: ...one level begins the programming sequence Key in 0 to 6 5535 1 0000 B3 Adjust Load Cell Corner 3 Factor This parameter allows you to adjust tweak the digital corner com pensation factory for load cell...

Page 15: ...s only 1 Switch off the force data instrument by pressing and holding the ON PRINT key for about 5 seconds 2 Press and hold down the ON PRINT key about 20 seconds until the screen shows F 3 Press the...

Page 16: ...the calibration was successful the display will show EndC1 momentarily followed by C 2 for the second calibration point 9 Repeat steps 7 8 for C2 thru C7 At the conclusion of C7 the instrument displa...

Page 17: ...r in the actual positive rated capacity value 3 After setting the exact value press the NET GROSS Set key to save the value The indicator will prompt you to enter the positive span in mV V P S 4 Use t...

Page 18: ...ollowing in sequence and then returns to the F CAL level Span Point 1 Weight Value Span Point 1 A D Count Span Point 2 Weight Value Span Point 2 A D Count Span Point 3 Weight Value Span Point 3 A D Co...

Page 19: ...on is transmitted continuously making it a popular choice for remote displays and other remote devices requiring a constant data stream The trans mission automatically occurs at the end of each displa...

Page 20: ...ed if it is disabled in A18 of the User Menu STRING FORMAT 1 Condec Demand String String Format 1 is designed for two way communication STX POL xxxxx xx LB KG GR NT CR LF Polarity SP Positive Negative...

Page 21: ...ogram Gross Net G Gross N Net Start Transmission Carriage Return Line Feed Status SP Valid M Motion O Over under range P pieces C custom units PCS pieces NOTE Additional unit is available N Newtons O...

Reviews: