background image

11 

 

Setting system time and date (A20)

 

 

1.  Switch off the instrument by pressing and holding down the ZERO/OFF key for about 5 sec-

onds. 

2.  Press  and  hold  down  the  ON/PRINT  key  (about  20  seconds)  until  the  screen  shows  

“-F-”. 

3.  Press the ON/PRINT (right) key once. The screen displays “-A-”. 

4.  Scroll down using the ZERO/OFF (down) key to reach the parameter level. The instrument 

shows “A   1”. 

5.  Move  from  A1  to  A20  by  pressing  the  TARE  (left)  key  repeatedly  until  the  screen  shows  

“A  20”. 

6.  Once you have arrived at A20 press the ZERO/OFF (down) key once. The screen displays 

“ho_xx” where ‘xx’ is the current hour, e.g. “15”. One digit will be flashing. 

7.  Use the four directional keys to adjust the displayed value to the actual hour value. Increase 

the  flashing  digit  by  pressing  the  UNITS  key.  Decrease  the  flashing  digit  by  pressing  the 
ZERO/OFF key.  Pressing  the  TARE key  or  the  ON/PRINT key  will change  the  position of 
the flashing digit. 

8.  After  entering  the  exact  value,  press  the  NET/GROSS  key  to  save  the  value.  The  screen 

displays “n¬_xx” where ‘xx’ is the current minute, e.g. “55”. One digit will be flashing.  

9.  Use  the  four  directional  keys  to  adjust  the  displayed  value  to  the  actual  minute  value.  In-

crease the flashing digit by pressing the UNITS key. Decrease the flashing digit by pressing 
the ZERO/OFF key. Pressing the TARE key or the ON/PRINT key will change the position 
of the flashing digit. 

10.  After  entering  the  exact  value,  press  the  NET/GROSS  key  to  save  the  value.  The  screen 

displays “dA_xx” where ‘xx’ is the current day of the month, e.g. “14”. One digit will be flash-
ing.  

11.  Use the four directional keys to adjust the displayed value to the actual day value. Increase 

the  flashing  digit  by  pressing  the  UNITS  key.  Decrease  the  flashing  digit  by  pressing  the 
ZERO/OFF key.  Pressing  the  TARE key  or  the  ON/PRINT key  will change  the  position of 
the flashing digit. 

12.  After  entering  the  exact  value,  press  the  NET/GROSS  key  to  save  the  value.  The  screen 

displays “n¬_xx” where ‘xx’ is the current month of the year, e.g. “02”. One digit will be flash-
ing. 

13.  Use  the  four  directional  keys  to  adjust  the  displayed  value  to  the  actual  month  value.  In-

crease the flashing digit by pressing the UNITS key. Decrease the flashing digit by pressing 
the ZERO/OFF key. Pressing the TARE key or the ON/PRINT key will change the position 
of the flashing digit. 

14.  After  entering  the  exact  value,  press  the  NET/GROSS  key  to  save  the  value.  The  screen 

displays “yE_xx” where ‘xx’ is the current month of the year, e.g. “11”. One digit will be flash-
ing. 

15.  Use the four directional keys to adjust the displayed value to the actual year value. Increase 

the  flashing  digit  by  pressing  the  UNITS  key.  Decrease  the  flashing  digit  by  pressing  the 
ZERO/OFF key.  Pressing  the  TARE key  or  the  ON/PRINT key  will change  the  position of 
the flashing digit. 

16.  After entering the exact value, press the NET/GROSS key to save the value and revert back 

up to the parameter level, e.g., “A   20”. 

Summary of Contents for TI-500 FDI Series

Page 1: ...MODEL TI 500 FDI Force Data Instrument Installer s Manual Revision 1 1 October 14 2016...

Page 2: ...OF TI 500 FORCE DATA INSTRUMENT 2 GETTING STARTED 3 SYSTEM CONFIGURATION 4 CONFIGURATION MENUS 4 ENTERING THE SETUP F CONFIGURATION MENU 5 SETUP F MENU DESCRIPTIONS 5 ENTERING THE USER A OR COM C MENU...

Page 3: ...libration of the force data instrument For operation and troubleshooting please refer to the separate user s guide Installation of TI 500 force data instrument Find a suitable location for the instrum...

Page 4: ...batteries to operate not included To install the batteries remove the plastic battery cover from the rear panel Observe proper direction polari ty of the batteries Replace the battery cover The TI 500...

Page 5: ...NOTE This menu will not appear for cabled systems see F29 The configuration menus are laid out in the following vertical arrangement Top Menu selection level Parameter level Selection level or functi...

Page 6: ...meter settings it is imperative to exit all menus in the following manner At the parameter level press the UNITS up key twice two times to exit the Setup Menu The instrument displays SAvE and then aut...

Page 7: ...re Press ZERO OFF key to begin se quence F16 Zero Calibra tion Places instrument into the zero calibration routine Scrolling down with the ZERO OFF key one level begins the procedure Press ZERO OFF ke...

Page 8: ...e AdC Internal A D cabled 1rAdIo One external wireless A D module 2rAdIo Two external wireless A D modules AdC 1rAdIo 2rAdIo F30 Special Appli cation Used to select one special application feature sub...

Page 9: ...o go from A1 to A2 press the ON PRINT key To go from A2 back to A1 press the TARE key 6 Once you have arrived at the proper A menu parameter e g A 1 press the ZERO OFF down key once to arrive at the s...

Page 10: ...bit 7_o_2 7 data bits with odd parity bit and two stop bits 8_n_1 8_n_2 7_n_1 7_n_2 7_E_1 7_E_2 7_o_1 7_o_2 A3 C3 Serial Port Mode Selects the mode of the serial port Refer to Appendix B for more info...

Page 11: ...lects the printed format for date USA mm dd yy IntL dd mm yy USA IntL A20 Set System Time Date Allows you to set the system time and date Pressing the ZERO OFF key to scroll down one level begins the...

Page 12: ...ashing digit 10 After entering the exact value press the NET GROSS key to save the value The screen displays dA_xx where xx is the current day of the month e g 14 One digit will be flash ing 11 Use th...

Page 13: ...uously out both serial ports TEST1 on COM1 and TEST2 on COM2 End test manually by pressing the NET GROSS Set key A24 U6 Keyboard Test Displays a keycode for each key pressed on the keypad See Table be...

Page 14: ...one level begins the programming sequence Key in 0 to 6 5535 1 0000 B3 Adjust Load Cell Corner 3 Factor This parameter allows you to adjust tweak the digital corner com pensation factory for load cell...

Page 15: ...s only 1 Switch off the force data instrument by pressing and holding the ON PRINT key for about 5 seconds 2 Press and hold down the ON PRINT key about 20 seconds until the screen shows F 3 Press the...

Page 16: ...the calibration was successful the display will show EndC1 momentarily followed by C 2 for the second calibration point 9 Repeat steps 7 8 for C2 thru C7 At the conclusion of C7 the instrument displa...

Page 17: ...r in the actual positive rated capacity value 3 After setting the exact value press the NET GROSS Set key to save the value The indicator will prompt you to enter the positive span in mV V P S 4 Use t...

Page 18: ...ollowing in sequence and then returns to the F CAL level Span Point 1 Weight Value Span Point 1 A D Count Span Point 2 Weight Value Span Point 2 A D Count Span Point 3 Weight Value Span Point 3 A D Co...

Page 19: ...on is transmitted continuously making it a popular choice for remote displays and other remote devices requiring a constant data stream The trans mission automatically occurs at the end of each displa...

Page 20: ...ed if it is disabled in A18 of the User Menu STRING FORMAT 1 Condec Demand String String Format 1 is designed for two way communication STX POL xxxxx xx LB KG GR NT CR LF Polarity SP Positive Negative...

Page 21: ...ogram Gross Net G Gross N Net Start Transmission Carriage Return Line Feed Status SP Valid M Motion O Over under range P pieces C custom units PCS pieces NOTE Additional unit is available N Newtons O...

Reviews: