10 

 

CODE/NAME 

DESCRIPTION 

SELECTION LIST 

A9/C9

 

Line Feeds

 

Allows  you  to  key-in  the  number  of  line  feeds.  Pressing  the  ZE-
RO/OFF key to scroll down one level begins the sequence. 

Key-in  
0 to 99 

8

 

A10

 

Auto Power Off 
- RF Digital 
Instrument 

 

Allows you to configure the automatic power off time for the RF force 
data instrument. Expressed in minutes of inactivity (keys and weigh-
ing  platform).  Pressing  the  ZERO/OFF  key  to  scroll  down  one  level 
begins the sequence. 

Key-in  
0 to 30 

30

 

A12

 

Backlight 
Brightness 

Selects  the  brightness  of  the  LCD  backlight.  Selections  are  in  %  of 
full brightness. 

0 (OFF) 
20 
50 
75 

100

 

A13/C13

 

Handshaking 

 

Selects function of the hardware handshaking. (NOTE: Receive pin is 
used for handshaking). 

"0" = Disabled 

"1" = Enabled 

0

 

A18

 

Date

 

Selects function of the printed date. 

"no" = Disabled 

"YES" = Enabled 

no

 

YES 

A19

 

Date Format

 

Selects the printed format for date.  

"USA" = mm/dd/yy       "IntL" = dd/mm/yy 

USA

      

IntL 

 

A20 

Set System 
Time & Date 

Allows you to set the system time and date. Pressing the ZERO/OFF 
key to scroll down one level begins the sequence. 

Press 

ZERO/OFF

 

key to begin se-
quence

 

A22

 

Low Battery 
Auto Power Off 
- RF Digital 
Instrument 

 

Allows  you  to  configure  the  automatic  power  off  time  of  the  instru-
ment  after  it  enters  a  low  battery  condition.  Expressed  in  minutes. 
Pressing the ZERO/OFF key to scroll down one level begins the se-
quence. 

Key-in  
0 to 99 

2

 

A23

 

Audible Key 
Feedback

 

Selects function of the audible key feedback (beeper). 

"no" = Disabled 

"YES" = Enabled 

no 

YES

 

A24

 

Diagnostics 

 

Used to access the listed test functions (one at a time). Pressing the 

ZERO/OFF

 key begins the sequence. 

"A24-U1" = Display segment  test, "A24-U2" = A/D converter test,  

“A24-U3” = Input test, “A24-U4” = Output test (all) 

“A24-U5” = Serial Port test (both), “A24-U6” = Keyboard test 

Press 

ZERO/OFF

 

key to begin se-
quence 

A25

 

Decimal Point 

 

Selects printed (not displayed) decimal point character.  

"0" = Period (‘.’) 

"1" = Comma (‘,’) 

 

1  

 

 

 

Summary of Contents for TI-500 FDI Series

Page 1: ...MODEL TI 500 FDI Force Data Instrument Installer s Manual Revision 1 1 October 14 2016...

Page 2: ...OF TI 500 FORCE DATA INSTRUMENT 2 GETTING STARTED 3 SYSTEM CONFIGURATION 4 CONFIGURATION MENUS 4 ENTERING THE SETUP F CONFIGURATION MENU 5 SETUP F MENU DESCRIPTIONS 5 ENTERING THE USER A OR COM C MENU...

Page 3: ...libration of the force data instrument For operation and troubleshooting please refer to the separate user s guide Installation of TI 500 force data instrument Find a suitable location for the instrum...

Page 4: ...batteries to operate not included To install the batteries remove the plastic battery cover from the rear panel Observe proper direction polari ty of the batteries Replace the battery cover The TI 500...

Page 5: ...NOTE This menu will not appear for cabled systems see F29 The configuration menus are laid out in the following vertical arrangement Top Menu selection level Parameter level Selection level or functi...

Page 6: ...meter settings it is imperative to exit all menus in the following manner At the parameter level press the UNITS up key twice two times to exit the Setup Menu The instrument displays SAvE and then aut...

Page 7: ...re Press ZERO OFF key to begin se quence F16 Zero Calibra tion Places instrument into the zero calibration routine Scrolling down with the ZERO OFF key one level begins the procedure Press ZERO OFF ke...

Page 8: ...e AdC Internal A D cabled 1rAdIo One external wireless A D module 2rAdIo Two external wireless A D modules AdC 1rAdIo 2rAdIo F30 Special Appli cation Used to select one special application feature sub...

Page 9: ...o go from A1 to A2 press the ON PRINT key To go from A2 back to A1 press the TARE key 6 Once you have arrived at the proper A menu parameter e g A 1 press the ZERO OFF down key once to arrive at the s...

Page 10: ...bit 7_o_2 7 data bits with odd parity bit and two stop bits 8_n_1 8_n_2 7_n_1 7_n_2 7_E_1 7_E_2 7_o_1 7_o_2 A3 C3 Serial Port Mode Selects the mode of the serial port Refer to Appendix B for more info...

Page 11: ...lects the printed format for date USA mm dd yy IntL dd mm yy USA IntL A20 Set System Time Date Allows you to set the system time and date Pressing the ZERO OFF key to scroll down one level begins the...

Page 12: ...ashing digit 10 After entering the exact value press the NET GROSS key to save the value The screen displays dA_xx where xx is the current day of the month e g 14 One digit will be flash ing 11 Use th...

Page 13: ...uously out both serial ports TEST1 on COM1 and TEST2 on COM2 End test manually by pressing the NET GROSS Set key A24 U6 Keyboard Test Displays a keycode for each key pressed on the keypad See Table be...

Page 14: ...one level begins the programming sequence Key in 0 to 6 5535 1 0000 B3 Adjust Load Cell Corner 3 Factor This parameter allows you to adjust tweak the digital corner com pensation factory for load cell...

Page 15: ...s only 1 Switch off the force data instrument by pressing and holding the ON PRINT key for about 5 seconds 2 Press and hold down the ON PRINT key about 20 seconds until the screen shows F 3 Press the...

Page 16: ...the calibration was successful the display will show EndC1 momentarily followed by C 2 for the second calibration point 9 Repeat steps 7 8 for C2 thru C7 At the conclusion of C7 the instrument displa...

Page 17: ...r in the actual positive rated capacity value 3 After setting the exact value press the NET GROSS Set key to save the value The indicator will prompt you to enter the positive span in mV V P S 4 Use t...

Page 18: ...ollowing in sequence and then returns to the F CAL level Span Point 1 Weight Value Span Point 1 A D Count Span Point 2 Weight Value Span Point 2 A D Count Span Point 3 Weight Value Span Point 3 A D Co...

Page 19: ...on is transmitted continuously making it a popular choice for remote displays and other remote devices requiring a constant data stream The trans mission automatically occurs at the end of each displa...

Page 20: ...ed if it is disabled in A18 of the User Menu STRING FORMAT 1 Condec Demand String String Format 1 is designed for two way communication STX POL xxxxx xx LB KG GR NT CR LF Polarity SP Positive Negative...

Page 21: ...ogram Gross Net G Gross N Net Start Transmission Carriage Return Line Feed Status SP Valid M Motion O Over under range P pieces C custom units PCS pieces NOTE Additional unit is available N Newtons O...

Reviews: